Correction of nonuniform attenuation and image fusion in SPECT imaging by means of separate X-ray CT
- PMID: 12126095
- DOI: 10.1007/BF03000104
Correction of nonuniform attenuation and image fusion in SPECT imaging by means of separate X-ray CT
Abstract
Improvements in image quality and quantitation measurement, and the addition of detailed anatomical structures are important topics for single-photon emission tomography (SPECT). The goal of this study was to develop a practical system enabling both nonuniform attenuation correction and image fusion of SPECT images by means of high-performance X-ray computed tomography (CT). A SPECT system and a helical X-ray CT system were placed next to each other and linked with Ethernet. To avoid positional differences between the SPECT and X-ray CT studies, identical flat patient tables were used for both scans; body distortion was minimized with laser beams from the upper and lateral directions to detect the position of the skin surface. For the raw projection data of SPECT, a scatter correction was performed with the triple energy window method. Image fusion of the X-ray CT and SPECT images was performed automatically by auto-registration of fiducial markers attached to the skin surface. After registration of the X-ray CT and SPECT images, an X-ray CT-derived attenuation map was created with the calibration curve for 99mTc. The SPECT images were then reconstructed with scatter and attenuation correction by means of a maximum likelihood expectation maximization algorithm. This system was evaluated in torso and cylindlical phantoms and in 4 patients referred for myocardial SPECT imaging with Tc-99m tetrofosmin. In the torso phantom study, the SPECT and X-ray CT images overlapped exactly on the computer display. After scatter and attenuation correction, the artifactual activity reduction in the inferior wall of the myocardium improved. Conversely, the incresed activity around the torso surface and the lungs was reduced. In the abdomen, the liver activity, which was originally uniform, had recovered after scatter and attenuation correction processing. The clinical study also showed good overlapping of cardiac and skin surface outlines on the fused SPECT and X-ray CT images. The effectiveness of the scatter and attenuation correction process was similar to that observed in the phantom study. Because the total time required for computer processing was less than 10 minutes, this method of attenuation correction and image fusion for SPECT images is expected to become popular in clinical practice.
Similar articles
-
Correction of photon attenuation and collimator response for a body-contouring SPECT/CT imaging system.J Nucl Med. 2005 May;46(5):868-77. J Nucl Med. 2005. PMID: 15872362 Clinical Trial.
-
Attenuation correction of myocardial SPECT images with X-ray CT: effects of registration errors between X-ray CT and SPECT.Ann Nucl Med. 2002 Sep;16(6):431-5. doi: 10.1007/BF02990083. Ann Nucl Med. 2002. PMID: 12416584
-
A study on attenuation correction using Tc-99m external TCT source in Tc-99m GSA liver SPECT.Ann Nucl Med. 2004 Oct;18(7):585-90. doi: 10.1007/BF02984580. Ann Nucl Med. 2004. PMID: 15586632 Clinical Trial.
-
X-ray-based attenuation correction for positron emission tomography/computed tomography scanners.Semin Nucl Med. 2003 Jul;33(3):166-79. doi: 10.1053/snuc.2003.127307. Semin Nucl Med. 2003. PMID: 12931319 Review.
-
Single-photon emission computed tomography/computed tomography: basic instrumentation and innovations.Semin Nucl Med. 2006 Oct;36(4):258-66. doi: 10.1053/j.semnuclmed.2006.05.005. Semin Nucl Med. 2006. PMID: 16950143 Review.
Cited by
-
Clinical validation of SPECT attenuation correction using x-ray computed tomography-derived attenuation maps: multicenter clinical trial with angiographic correlation.J Nucl Cardiol. 2005 Nov-Dec;12(6):676-86. doi: 10.1016/j.nuclcard.2005.08.006. J Nucl Cardiol. 2005. PMID: 16344230 Clinical Trial.
-
Measuring hepatic functional reserve using low temporal resolution Gd-EOB-DTPA dynamic contrast-enhanced MRI: a preliminary study comparing galactosyl human serum albumin scintigraphy with indocyanine green retention.Eur Radiol. 2014 Jan;24(1):112-9. doi: 10.1007/s00330-013-2983-y. Epub 2013 Aug 15. Eur Radiol. 2014. PMID: 23949726
-
Simultaneous Emission-Transmission Tomography in an MRI Hardware Framework.IEEE Trans Radiat Plasma Med Sci. 2018 Jul;2(4):326-336. doi: 10.1109/TRPMS.2018.2835312. Epub 2018 May 10. IEEE Trans Radiat Plasma Med Sci. 2018. PMID: 29998213 Free PMC article.
-
On the Effect of the Patient Table on Attenuation in Myocardial Perfusion Imaging SPECT.EJNMMI Phys. 2025 Jan 20;12(1):3. doi: 10.1186/s40658-024-00713-4. EJNMMI Phys. 2025. PMID: 39832088 Free PMC article.
-
Simplified quantification method for in vivo SPECT/CT imaging of asialoglycoprotein receptor with (99m)Tc-p(VLA-co-VNI) to assess and stage hepatic fibrosis in mice.Sci Rep. 2016 May 6;6:25377. doi: 10.1038/srep25377. Sci Rep. 2016. PMID: 27150943 Free PMC article.