Noninvasive reduction of open-book pelvic fractures by circumferential compression

Affiliations

Noninvasive reduction of open-book pelvic fractures by circumferential compression

Michael Bottlang et al. J Orthop Trauma. 2002 Jul.

Abstract

Objectives: To determine the efficacy and optimal application parameters of circumferential compression to reduce external rotation-type pelvic fractures.

Design: Biomechanical investigation on human cadaveric specimens.

Setting: Biomechanics laboratory.

Intervention: Partially stable and unstable external rotation injuries of the pelvic ring (OTA classification 61-B1 and 61-C1) were created in seven human cadaveric specimens. A prototype pelvic strap was applied subsequently at three distinct transverse levels around the pelvis. Circumferential pelvic compression was induced by gradual tensioning of the strap to attempt complete reduction of the symphysis diastasis.

Main outcome measurements: Pelvic reduction was evaluated with respect to strap tension and the strap application site. The effect of circumferential compression on intraperitoneal pressure and skin-strap interface pressure was measured.

Results: A successive increase in circumferential compression consistently induced a gradual decrease in symphysis diastasis. An optimal strap application site was determined, at which circumferential compression most effectively yielded pelvic reduction. The minimum strap tension required to achieve complete reduction of symphysis diastasis was determined to be 177 +/- 44 Newtons and 180 +/- 50 Newtons in the partially stable and unstable pelvis, respectively.

Conclusions: Application of circumferential compression to the pelvic soft tissue envelope with a pelvic strap was an efficient means to achieve controlled reduction of external rotation-type pelvic fractures. This study derived application parameters with direct clinical implication for noninvasive emergent management of traumatic pelvic ring disruptions.