Properties of perimetric threshold estimates from Full Threshold, SITA Standard, and SITA Fast strategies
- PMID: 12147599
Properties of perimetric threshold estimates from Full Threshold, SITA Standard, and SITA Fast strategies
Abstract
Purpose: To investigate the distributions of threshold estimates with the Swedish Interactive Threshold Algorithms (SITA) Standard, SITA Fast, and the Full Threshold algorithm (Humphrey Field Analyzer; Zeiss-Humphrey Instruments, Dublin, CA) and to compare the pointwise test-retest variability of these strategies.
Methods: One eye of 49 patients (mean age, 61.6 years; range, 22-81) with glaucoma (Mean Deviation mean, -7.13 dB; range, +1.8 to -23.9 dB) was examined four times with each of the three strategies. The mean and median SITA Standard and SITA Fast threshold estimates were compared with a "best available" estimate of sensitivity (mean results of three Full Threshold tests). Pointwise 90% retest limits (5th and 95th percentiles of retest thresholds) were derived to assess the reproducibility of individual threshold estimates.
Results: The differences between the threshold estimates of the SITA and Full Threshold strategies were largest ( approximately 3 dB) for midrange sensitivities ( approximately 15 dB). The threshold distributions of SITA were considerably different from those of the Full Threshold strategy. The differences remained of similar magnitude when the analysis was repeated on a subset of 20 locations that are examined early during the course of a Full Threshold examination. With sensitivities above 25 dB, both SITA strategies exhibited lower test-retest variability than the Full Threshold strategy. Below 25 dB, the retest intervals of SITA Standard were slightly smaller than those of the Full Threshold strategy, whereas those of SITA Fast were larger.
Conclusions: SITA Standard may be superior to the Full Threshold strategy for monitoring patients with visual field loss. The greater test-retest variability of SITA Fast in areas of low sensitivity is likely to offset the benefit of even shorter test durations with this strategy. The sensitivity differences between the SITA and Full Threshold strategies may relate to factors other than reduced fatigue. They are, however, small in comparison to the test-retest variability.
Similar articles
-
Threshold and variability properties of matrix frequency-doubling technology and standard automated perimetry in glaucoma.Invest Ophthalmol Vis Sci. 2005 Jul;46(7):2451-7. doi: 10.1167/iovs.05-0135. Invest Ophthalmol Vis Sci. 2005. PMID: 15980235
-
SITA standard in optic neuropathies and hemianopias: a comparison with full threshold testing.Invest Ophthalmol Vis Sci. 2001 Feb;42(2):528-37. Invest Ophthalmol Vis Sci. 2001. PMID: 11157893
-
The SITA perimetric threshold algorithms in glaucoma.Invest Ophthalmol Vis Sci. 1999 Aug;40(9):1998-2009. Invest Ophthalmol Vis Sci. 1999. PMID: 10440254
-
Selective perimetry in glaucoma diagnosis.Curr Opin Ophthalmol. 2007 Mar;18(2):115-21. doi: 10.1097/ICU.0b013e3280555096. Curr Opin Ophthalmol. 2007. PMID: 17301612 Review.
-
Visual field testing in glaucoma using the Swedish Interactive Thresholding Algorithm (SITA).Surv Ophthalmol. 2025 Jan-Feb;70(1):141-152. doi: 10.1016/j.survophthal.2024.09.005. Epub 2024 Sep 29. Surv Ophthalmol. 2025. PMID: 39349186 Review.
Cited by
-
Retinal Ganglion Cell Content Underlying Standard Automated Perimetry Size I to V Visual Sensitivities in the Non-Human Primate Experimental Glaucoma Model.Invest Ophthalmol Vis Sci. 2024 Jul 1;65(8):22. doi: 10.1167/iovs.65.8.22. Invest Ophthalmol Vis Sci. 2024. PMID: 38995114 Free PMC article.
-
Imaging Glaucomatous Damage Across the Temporal Raphe.Invest Ophthalmol Vis Sci. 2015 Jun;56(6):3496-504. doi: 10.1167/iovs.15-16730. Invest Ophthalmol Vis Sci. 2015. PMID: 26047040 Free PMC article.
-
An Artificial Intelligence Approach to Detect Visual Field Progression in Glaucoma Based on Spatial Pattern Analysis.Invest Ophthalmol Vis Sci. 2019 Jan 2;60(1):365-375. doi: 10.1167/iovs.18-25568. Invest Ophthalmol Vis Sci. 2019. PMID: 30682206 Free PMC article.
-
Choice of Stimulus Range and Size Can Reduce Test-Retest Variability in Glaucomatous Visual Field Defects.Transl Vis Sci Technol. 2014 Sep 25;3(5):6. doi: 10.1167/tvst.3.5.6. eCollection 2014 Sep. Transl Vis Sci Technol. 2014. PMID: 25371855 Free PMC article.
-
Homonymous hemianopia: challenges and solutions.Clin Ophthalmol. 2014 Sep 22;8:1919-27. doi: 10.2147/OPTH.S59452. eCollection 2014. Clin Ophthalmol. 2014. PMID: 25284978 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous