[Detection of left ventricular endocardial border with tissue harmonic imaging]
- PMID: 12147967
[Detection of left ventricular endocardial border with tissue harmonic imaging]
Abstract
Background: Ultrasonic wave propagation in human tissues is not linear. As a consequence, harmonic waves, whose frequency is a multiple (harmonic) of the emitted frequency, are generated. Tissue Harmonic Imaging (THI) processes only the second harmonic frequency in order to improve the signal-to-noise ratio of the received signal. The study was aimed at investigating the impact of THI on the detection of the Left Ventricular (LV) endocardial border. Attention was paid to determine which LV walls were analysed more clearly with THI rather than with conventional Fundamental Frequency Imaging (FFI).
Methods: We compared the FFI and the THI visualization of the 16 LV segments and of the 6 LV walls in 30 consecutive patients by scoring the images as bad, sufficient or good. The equipment used was a GE Vingmed System Five (Orten, Norway).
Results: Images were good in 85% of segments explored with THI, whereas FFI provided good images in 47% of segments (p<0.001). From the apical window, visualization of the apex, lateral wall and anterior wall significantly improved with THI; using the parasternal approach, imaging of posterior wall was definitely better with THI than with FFI.
Conclusions: THI enhances both endocardial visualization and global image quality. Tech-nical development, however, increases the diagnostic possibilities of echocardiography but does not substitute the operator's experience.
Similar articles
-
Pulse inversion harmonic imaging improves endocardial border visualization in two-dimensional images: comparison with harmonic imaging.J Am Soc Echocardiogr. 2002 Apr;15(4):302-8. doi: 10.1067/mje.2002.117294. J Am Soc Echocardiogr. 2002. PMID: 11944006
-
Comparison of native and contrast-enhanced harmonic echocardiography for visualization of left ventricular endocardial border.Am J Cardiol. 1999 Jan 15;83(2):211-7. doi: 10.1016/s0002-9149(98)00826-1. Am J Cardiol. 1999. PMID: 10073823 Clinical Trial.
-
Contrast-enhanced harmonic color Doppler for left ventricular opacification: improved endocardial border definition compared to tissue harmonic imaging and optimization of methodologyin patients with suboptimal echocardiograms.Echocardiography. 2001 Nov;18(8):639-49. doi: 10.1046/j.1540-8175.2001.00639.x. Echocardiography. 2001. PMID: 11801205
-
Tissue harmonic imaging for standard left ventricular measurements: fundamentally flawed?Eur J Echocardiogr. 2006 Jan;7(1):9-15. doi: 10.1016/j.euje.2005.08.006. Epub 2005 Oct 10. Eur J Echocardiogr. 2006. PMID: 16219493 Review.
-
[Improved endocardium imaging using modified transthoracic echocardiography with the second harmonic frequency (tissue harmonic imaging)].Herz. 1998 Dec;23(8):467-73. doi: 10.1007/BF03043753. Herz. 1998. PMID: 10023580 Review. German.