Calculated absorbed-dose ratios, TG51/TG21, for most widely used cylindrical and parallel-plate ion chambers over a range of photon and electron energies
- PMID: 12148727
- DOI: 10.1118/1.1487857
Calculated absorbed-dose ratios, TG51/TG21, for most widely used cylindrical and parallel-plate ion chambers over a range of photon and electron energies
Abstract
Task Group 51 (TG51), of the Radiation Therapy Committee of the American Association of Physicists in Medicine (AAPM), has developed a calibration protocol for high-energy photon and electron therapy beams based on absorbed dose standards. This protocol is intended to replace the air-kerma based protocol developed by an earlier AAPM task group (TG21). Conversion to the newer protocol introduces a change in the determined absorbed dose. In this work, the change in dose is expressed as the ratio of the doses (TG51/TG21) based on the two protocols. Dose is compared at the TG-51 reference depths of 10 cm for photons and d(ref) for electrons. Dose ratios are presented for a variety of ion chambers over a range of photon and electron energies. The TG51/TG21 dose ratios presented here are based on the dosimetry factors provided by the two protocols and the chamber-specific absorbed dose and exposure calibration factors (N60Co(D,w) and Nx) provided by the Accredited Dosimetry Calibration Laboratory (ADCL) at The University of Texas, M. D. Anderson Cancer Center (MDACC). As such, the values presented here represent the expected discrepancies between the two protocols due only to changes in the dosimetry parameters and the differences in chamber-specific dose and air-kerma standards. These values are independent of factors such as measurement uncertainties, setup errors, and inconsistencies arising from the mix of different phantoms and ion chambers for the two protocols. Therefore, these ratios may serve as a guide for institutions performing measurements for the switch from TG21-to-TG51 based calibration. Any significant deviation in the ratio obtained from measurements versus those presented here should prompt a review to identify possible errors and inconsistencies. For all cylindrical chambers included here, the TG51/TG21 dose ratios are the same within +/-0.6%, irrespective of the make and model of the chamber, for each photon and electron beam included. Photon beams show the TG51/TG21 dose ratios decreasing with energy, whereas electrons exhibit the opposite trend. The dose ratio for photons is near 1.00 at 18 mV increasing to near 1.01 at 4 mV while the dose ratio for electrons is near 1.02 at 20 MeV decreasing only 0.5% to near 1.015 at 6 MeV. For parallel-plate chambers, the situation is complicated by the two possible methods of obtaining calibration factors: through an ADCL or through a cross-comparison with a cylindrical chamber in a high-energy electron beam. For some chambers, the two methods lead to significantly different calibration factors, which in turn lead to significantly different TG51/TG21 results for the same chamber. Data show that if both N60Co(D,w) and Nx are obtained from the same source, namely an ADCL or a cross comparison, the TG51/TG21 results for parallel-plate chambers are similar to those for cylindrical chambers. However, an inconsistent set of calibration factors, i.e., using N60Co(D,w) x k(ecal) from an ADCL but Ngas from a cross comparison or vice versa, can introduce an additional uncertainty up to 2.5% in the TG51/TG21 dose ratios.
Comment in
-
Comment on "Calculated absorbed-dose ratios, TG51/TG21, for most widely used cylindrical and parallel-plate ion chambers over a range of photon and electron energies".Med Phys. 2003 Mar;30(3):473-7; author reply 478-80. doi: 10.1118/1.1541251. Med Phys. 2003. PMID: 12674249 No abstract available.
Similar articles
-
Reference dosimetry in clinical high-energy electron beams: comparison of the AAPM TG-51 and AAPM TG-21 dosimetry protocols.Med Phys. 2001 Oct;28(10):2077-87. doi: 10.1118/1.1405841. Med Phys. 2001. PMID: 11695769
-
Consistency of absorbed dose to water measurements using 21 ion-chamber models following the AAPM TG51 and TG21 calibration protocols.Med Phys. 2006 Jun;33(6):1818-28. doi: 10.1118/1.2199598. Med Phys. 2006. PMID: 16872089
-
Comparison of the IAEA TRS-398 and AAPM TG-51 absorbed dose to water protocols in the dosimetry of high-energy photon and electron beams.Phys Med Biol. 2001 Nov;46(11):2985-3006. doi: 10.1088/0031-9155/46/11/315. Phys Med Biol. 2001. PMID: 11720359
-
Advances in the determination of absorbed dose to water in clinical high-energy photon and electron beams using ionization chambers.Phys Med Biol. 2004 Feb 21;49(4):R49-104. doi: 10.1088/0031-9155/49/4/r01. Phys Med Biol. 2004. PMID: 15005158 Review.
-
The protocols and codes of practice used for the determination of absorbed dose in megavoltage photon and electron beams.Crit Rev Biomed Eng. 1991;19(2-3):147-80. Crit Rev Biomed Eng. 1991. PMID: 1769240 Review.
Cited by
-
Differences in electron beam dosimetry using two commercial ionization chambers and the TG-21 protocol: another reason to switch to TG-51.J Appl Clin Med Phys. 2003 Spring;4(2):124-31. doi: 10.1120/jacmp.v4i2.2527. J Appl Clin Med Phys. 2003. PMID: 12777147 Free PMC article.
-
Addendum to the AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon beams.Med Phys. 2014 Apr;41(4):041501. doi: 10.1118/1.4866223. Med Phys. 2014. PMID: 24694120 Free PMC article.
-
TG-51: experience from 150 institutions, common errors, and helpful hints.J Appl Clin Med Phys. 2003 Spring;4(2):102-11. doi: 10.1120/jacmp.v4i2.2524. J Appl Clin Med Phys. 2003. PMID: 12777144 Free PMC article.
-
Radiation therapy for deep periocular cancer treatments when protons are unavailable: is combining electrons and orthovoltage therapy beneficial?J Radiat Res. 2018 Sep 1;59(5):593-603. doi: 10.1093/jrr/rry045. J Radiat Res. 2018. PMID: 30053071 Free PMC article.
-
Validating kQ =1.0 assumption in TG51 with PTW 30013 farmer chamber for Varian TrueBeam's 2.5 MV imaging beam.J Appl Clin Med Phys. 2018 May;19(3):351-354. doi: 10.1002/acm2.12290. Epub 2018 Mar 1. J Appl Clin Med Phys. 2018. PMID: 29493079 Free PMC article.
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Medical
Research Materials
Miscellaneous