Differential effect of 180 degrees and 360 degrees acquisition orbits on the accuracy of SPECT imaging: quantitative evaluation in phantoms
- PMID: 12163639
Differential effect of 180 degrees and 360 degrees acquisition orbits on the accuracy of SPECT imaging: quantitative evaluation in phantoms
Abstract
A circular 180 degrees acquisition orbit is considered standard for cardiac SPECT imaging. Theoretically, a 360 degrees acquisition orbit is preferred because of more complete Fourier spectral information on projection data. The differential effect of 180 degrees and 360 degrees acquisition orbits on image accuracy (homogeneity and defect size) was assessed quantitatively in phantom studies.
Methods: SPECT imaging with a dual-head gamma camera was performed on normal cardiac phantoms filled with a (99m)Tc solution, using 180 degrees and 360 degrees circular acquisition orbits. The phantoms were placed in the center of the orbit and at 5, 10, and 15 cm off center. Fillable defect inserts of different sizes were placed in the phantom to simulate myocardial perfusion defects. The homogeneity of count distribution in short-axis slices of the normal phantom was analyzed as the percentage of variability. Defects were quantified as a percentage of the entire phantom volume using circumferential count profiles and normal reference profiles.
Results: When normal phantoms were placed in the center of the orbit, percentage variability was not different whether a 180 degrees or 360 degrees acquisition orbit was used (4.2% +/- 0.1% vs. 4.4% +/- 0.2%, P = not statistically significant). However when normal phantoms were placed off center, SPECT imaging with a 180 degrees acquisition orbit showed increasing inhomogeneity, both visually and quantitatively (e.g., percentage variability for the 15-cm off-center position was 10.8% +/- 0.1% (P < 0.0001). SPECT imaging with a 360 degrees acquisition orbit showed similar homogeneity visually and quantitatively, whether the phantom was placed in or off the center (e.g., percentage variability for the 15-cm off-center position was 4.6% +/- 0.5%, P = not statistically significant). Quantification of phantom defects acquired with a 180 degrees orbit showed increasing overestimation of defect sizes with increasingly off-center positions. Quantification of phantom defects acquired with a 360 degrees orbit showed no effect from progressively off-center positions, although phantom defect sizes were mildly underestimated.
Conclusion: SPECT images acquired with a 180 degrees orbit may have significant erroneous inhomogeneity and overestimation of defect size, in particular when the target object is off the center of the orbit, as is commonly seen in clinical cardiac imaging. In contrast, SPECT images acquired with a 360 degrees orbit may provide more accurate quantitative information.
Comment in
-
Effect of acquisition orbit on SPECT in phantoms.J Nucl Med. 2003 Jun;44(6):993-5; author reply 995-6. J Nucl Med. 2003. PMID: 12791832 No abstract available.
-
Cardiac SPECT: 360 degrees circular acquisition may resolve defects of 180 degrees data.J Nucl Med. 2003 Jun;44(6):995; author reply 995-6. J Nucl Med. 2003. PMID: 12791833 No abstract available.
-
False-positive findings on myocardial perfusion SPECT.J Nucl Med. 2004 Sep;45(9):1597. J Nucl Med. 2004. PMID: 15347730 No abstract available.
Similar articles
-
Assessment of infarct size and severity by quantitative myocardial SPECT: results from a multicenter study using a cardiac phantom.J Nucl Med. 2000 Aug;41(8):1383-90. J Nucl Med. 2000. PMID: 10945532
-
Body-contour versus circular orbit acquisition in cardiac SPECT: assessment of defect detectability with channelized Hotelling observer.Nucl Med Commun. 2007 Dec;28(12):937-42. doi: 10.1097/MNM.0b013e3282f1b9a3. Nucl Med Commun. 2007. PMID: 18090221
-
Validation of the Yale circumferential quantification method using 201Tl and 99mTc: a phantom study.J Nucl Med. 2000 Aug;41(8):1436-41. J Nucl Med. 2000. PMID: 10945538
-
IQ·SPECT technology and its clinical applications using multicenter normal databases.Ann Nucl Med. 2017 Nov;31(9):649-659. doi: 10.1007/s12149-017-1210-3. Epub 2017 Sep 22. Ann Nucl Med. 2017. PMID: 28940141 Free PMC article. Review.
-
Quantitative imaging: systematic review of perfusion/flow phantoms.Eur Radiol Exp. 2020 Mar 4;4(1):15. doi: 10.1186/s41747-019-0133-2. Eur Radiol Exp. 2020. PMID: 32128653 Free PMC article.
Cited by
-
Performance of Myocardial Perfusion Imaging Using Multi-focus Fan Beam Collimator with Resolution Recovery Reconstruction in a Comparison with Conventional SPECT.Asia Ocean J Nucl Med Biol. 2014 Fall;2(2):111-9. Asia Ocean J Nucl Med Biol. 2014. PMID: 27408867 Free PMC article.
-
Comparison of 180 degrees and 360 degrees acquisition for myocardial perfusion SPECT with compensation for attenuation, detector response, and scatter: Monte Carlo and mathematical observer results.J Nucl Cardiol. 2006 May-Jun;13(3):345-53. doi: 10.1016/j.nuclcard.2006.03.008. J Nucl Cardiol. 2006. PMID: 16750779
-
Massive left pleural effusion resulting in false-positive myocardial perfusion.J Nucl Cardiol. 2005 Jul-Aug;12(4):476-9. doi: 10.1016/j.nuclcard.2005.03.001. J Nucl Cardiol. 2005. PMID: 16084438 No abstract available.
-
On Acquisition Parameters and Processing Techniques for Interparticle Contact Detection in Granular Packings Using Synchrotron Computed Tomography.J Imaging. 2022 May 12;8(5):135. doi: 10.3390/jimaging8050135. J Imaging. 2022. PMID: 35621899 Free PMC article.
-
Fast technetium-99m liver SPECT for evaluation of the pretreatment procedure for radioembolization dosimetry.Med Phys. 2019 Jan;46(1):345-355. doi: 10.1002/mp.13253. Epub 2018 Nov 13. Med Phys. 2019. PMID: 30347130 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources