Differential effect of right and left carotid sinus massage: implications for sudden rate drop pacing algorithm
- PMID: 12167966
Differential effect of right and left carotid sinus massage: implications for sudden rate drop pacing algorithm
Abstract
This present report describes a patient with syncope in whom asystole was triggered by right carotid sinus massage and complete atrioventricular block with slight bradycardia by left carotid sinus massage. A dual-chamber, rate-responsive, permanent pacemaker with 'sudden rate drop' algorithm was implanted. The present report underlines the pathophysiology and the clinical utility of carotid sinus massage on both sides. Programming considerations to tailor the rate drop algorithm when two different physiological responses are triggered by carotid sinus massage are discussed. The importance of careful clinical assessment to detect all appropriate physiological responses as well as new pacing algorithms is depicted.
Similar articles
-
Clinical experience of a new rate drop response algorithm in the treatment of vasovagal and carotid sinus syncope.Europace. 2000 Jul;2(3):245-50. doi: 10.1053/eupc.2000.0107. Europace. 2000. PMID: 11227596
-
[Analysis of programmed pacing rate levels in patients with the carotid sinus hypersensitivity syndrome and implanted Clarity pacemakers].Srp Arh Celok Lek. 2001 Jan-Feb;129(1-2):9-12. Srp Arh Celok Lek. 2001. PMID: 11534283 Serbian.
-
Sudden rate drop intervention rate level in patients with new 'clarity' pacemakers and carotid sinus syndrome.Med Sci Monit. 2001 Jan-Feb;7(1):64-7. Med Sci Monit. 2001. PMID: 11208495
-
Clinically oriented device programming in bradycardia patients: part 2 (atrioventricular blocks and neurally mediated syncope). Proposals from AIAC (Italian Association of Arrhythmology and Cardiac Pacing).J Cardiovasc Med (Hagerstown). 2018 Apr;19(4):170-180. doi: 10.2459/JCM.0000000000000629. J Cardiovasc Med (Hagerstown). 2018. PMID: 29470251 Review.
-
[Clinical cardiac electrophysiological studies in carotid sinus syndrome].Orv Hetil. 1994 Aug 7;135(32):1743-50. Orv Hetil. 1994. PMID: 8072748 Review. Hungarian.
Publication types
MeSH terms
LinkOut - more resources
Medical