Dosimetric characteristics of Novalis shaped beam surgery unit
- PMID: 12201420
- DOI: 10.1118/1.1494830
Dosimetric characteristics of Novalis shaped beam surgery unit
Abstract
The dosimetric characteristics of a new dedicated radiosurgical treatment unit are systematically measured in terms of its percent depth dose, beam profile, and relative scatter factor. High-resolution diode detector, mini-ion-chamber detector, and conventional Kodak XV films are used to measure dosimetric data for a range of field sizes from 6x6 mm to 100x100 mm. The effects of collimator size, micro-multileaf collimator shape, and detector type on the dosimetric data are investigated. Results indicate that, with careful design, accurate dosimetric data could be acquired using either a dedicated diode detector or a mini-ion-chamber detector, and film detector. Special attention is required when measuring dosimetric data for small field sizes such as 6x6 mm.
Similar articles
-
A new approach in dose measurement and error analysis for narrow photon beams (beamlets) shaped by different multileaf collimators using a small detector.Med Phys. 2004 Jul;31(7):2020-32. doi: 10.1118/1.1760191. Med Phys. 2004. PMID: 15305454
-
Dosimetric characteristics of novalis Tx system with high definition multileaf collimator.Med Phys. 2008 Oct;35(10):4460-3. doi: 10.1118/1.2977668. Med Phys. 2008. PMID: 18975693
-
Lateral loss and dose discrepancies of multileaf collimator segments in intensity modulated radiation therapy.Med Phys. 2003 Nov;30(11):2959-68. doi: 10.1118/1.1619230. Med Phys. 2003. PMID: 14655943
-
[Technologic developments in radiotherapy and stereotactic radiosurgery].Neurochirurgie. 2001 May;47(2-3 Pt 2):260-6. Neurochirurgie. 2001. PMID: 11404704 Review. French.
-
[Symposium: an attempt at radiation therapy using advanced techniques. 3. A characteristic and technological base of carbon beam therapy].Nihon Hoshasen Gijutsu Gakkai Zasshi. 2005 Jul 20;61(7):926-8. doi: 10.6009/jjrt.kj00003326891. Nihon Hoshasen Gijutsu Gakkai Zasshi. 2005. PMID: 16075489 Review. Japanese. No abstract available.
Cited by
-
Radiosurgery technology development and use.J Radiosurg SBRT. 2011;1(1):21-29. J Radiosurg SBRT. 2011. PMID: 29296294 Free PMC article.
-
Dosimetric and radiobiological comparison of treatment plan between CyberKnife and EDGE in stereotactic body radiotherapy for pancreatic cancer.Sci Rep. 2021 Feb 18;11(1):4065. doi: 10.1038/s41598-021-83648-5. Sci Rep. 2021. PMID: 33603030 Free PMC article.
-
Contemporary treatment with radiosurgery for spine metastasis and spinal cord compression in 2015.Radiat Oncol J. 2015 Mar;33(1):1-11. doi: 10.3857/roj.2015.33.1.1. Epub 2015 Mar 31. Radiat Oncol J. 2015. PMID: 25874172 Free PMC article. Review.
-
Physical and dosimetric characteristic of high-definition multileaf collimator (HDMLC) for SRS and IMRT.J Appl Clin Med Phys. 2011 Apr 14;12(3):3475. doi: 10.1120/jacmp.v12i3.3475. J Appl Clin Med Phys. 2011. PMID: 21844860 Free PMC article.
-
AAPM-RSS Medical Physics Practice Guideline 9.a. for SRS-SBRT.J Appl Clin Med Phys. 2017 Sep;18(5):10-21. doi: 10.1002/acm2.12146. Epub 2017 Aug 8. J Appl Clin Med Phys. 2017. PMID: 28786239 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources