[Radiation exposure of health personnel and patients in the heart catheterization laboratory in during vascular brachytherapy]
- PMID: 12219698
- DOI: 10.1007/s00392-002-0825-z
[Radiation exposure of health personnel and patients in the heart catheterization laboratory in during vascular brachytherapy]
Abstract
Interventional radiological measures can lead to high radiation exposures for medical staff. In order to determine the radiation exposure to staff and patients, the resulting radiation exposures were directly measured for 52 measures at an cardiac catheterization laboratory with a new dosimetry system DIS (Direct Ion Storage). Beside the measurement of body dose behind the lead apron, measurements of radiation doses were performed in front of the lead apron and at the wrist of the physician. These measurements were taken as an approximation of the radiation exposure of the non-shielded body parts. The patients dose was estimated by placing a dosemeter close to the head of the patient and from the dose-area product. The mean value of body dose from 52 measurements for the physician behind the lead apron was 1.9 microSv per procedure with a range of 0-9 microSv. In front of the lead apron, a mean value of 53.9 microSv (3-233 microSv) per procedure was obtained. The mean value of partial body dose at the physician's wrist was determined to be 163.2 microSv (12-603 microSv) per procedure. It could be shown that measures combined with interventions lead to higher exposures compared with measures without interventions. For the medical technician, the mean value behind the lead apron was 3.9 microSv (0-58 microSv) per procedure. For the patient, a mean value of 800 microSv (119-8642 microSv) was measured close to the head. The mean dose to the skin of the patient at radiation entrance was determined to be 307 mGy (70-1190 mGy). From this data, the radiation dose per year for the physician performing 1000 measures, was determined to be 1.9 mSv/year. This is below the new dose limit of 20 mSv/year. Also the estimations for the dose to the eye lens and the hands of the physician show no conflicts with actual dose limits when obeying all radiation regulations. The results for vascular brachytherapy did not show significantly higher exposures, compared with conventional measures including interventions.
Similar articles
-
Estimation of staff lens doses during interventional procedures. Comparing cardiology, neuroradiology and interventional radiology.Radiat Prot Dosimetry. 2015 Jul;165(1-4):279-83. doi: 10.1093/rpd/ncv049. Epub 2015 Apr 5. Radiat Prot Dosimetry. 2015. PMID: 25848117
-
[Estimation of radiation exposure and radiation risk for employees of a heart catheterization laboratory].Z Kardiol. 1997 Apr;86(4):258-63. doi: 10.1007/s003920050057. Z Kardiol. 1997. PMID: 9235797 German.
-
Radiation exposure of medical staff from interventional x-ray procedures: a multicentre study.Eur Radiol. 2009 Aug;19(8):2000-8. doi: 10.1007/s00330-009-1388-4. Epub 2009 Apr 7. Eur Radiol. 2009. PMID: 19350250
-
Occupational radiation doses to operators performing cardiac catheterization procedures.Health Phys. 2008 Mar;94(3):211-27. doi: 10.1097/01.HP.0000290614.76386.35. Health Phys. 2008. PMID: 18301095 Review.
-
Intracoronary brachytherapy in the Cath Lab. Physics dosimetry, technology and safety considerations.Herz. 1998 Sep;23(6):401-6. doi: 10.1007/BF03043606. Herz. 1998. PMID: 9816527 Review.
Cited by
-
Angiographic and clinical outcome for the treatment of in-stent restenosis with sirolimus-eluting stent compared to vascular brachytherapy.Z Kardiol. 2005 Jun;94(6):405-10. doi: 10.1007/s00392-005-0253-y. Z Kardiol. 2005. PMID: 15940441 Clinical Trial.
-
Occupational exposure to low doses of ionizing radiation and cataract development: a systematic literature review and perspectives on future studies.Radiat Environ Biophys. 2013 Aug;52(3):303-19. doi: 10.1007/s00411-013-0477-6. Epub 2013 Jun 27. Radiat Environ Biophys. 2013. PMID: 23807741
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical