[Image quality and detection of pathology by ultrasound: comparison of B-mode ultrasound with photopic imaging and tissue harmonic imaging alone and in combination]
- PMID: 12375209
- DOI: 10.1055/s-2002-34551
[Image quality and detection of pathology by ultrasound: comparison of B-mode ultrasound with photopic imaging and tissue harmonic imaging alone and in combination]
Abstract
Purpose: To determine the accuracy of photopic imaging (PI) in detecting pathology by ultrasound (US) and to assess the image quality in direct comparison with conventional B-mode ultrasound and tissue harmonic imaging (THI).
Material and methods: Fifty-two patients underwent US examination, among them 29 patients for abdominal assessment and 23 for otolaryngological assessment. A total of 208 freeze frames, 52 B-mode scans each with and without THI and 52 B-mode scans each with and without PI, were assessed by three readers, who determined the presence of pathology on a scale of 1 (definitely abnormal) to 5 (definitely normal). All 52 patients underwent US follow-up within six weeks. The results were confirmed by CT in 30 patients and by histology in five cases. Image quality and different color encodings of each technique were rated on a ranking scale of 1 (optimal) to 4 (poor). The different US techniques were compared in terms of image quality, diagnostic accuracy, and color encoding using McNemar's test and ROC analysis.
Results: The results for image quality were as follows: B-scan 3.9; THI 1.9; PI 2.8; and THI plus PI 1.5 (each p < 0.05). The following AUCs (Area under Curve, presence of pathology) were calculated: 0.925, 0.990, and 0.990 for B-mode US, THI, and PI, respectively (not significant), and 0.994 for THI plus PI (significant compared to B-mode scan). The different color encodings were rated as follows: reddish brown 1.6, gray 1.9, blue 3.1, and green 3.6 (each p < 0.05).
Conclusion: For ultrasound examinations, PI in combination with THI improves the image quality and conspicuity of pathology.
Similar articles
-
[Image quality and characterisation of sonographically detected breast lesions: comparison of tissue harmonic imaging alone and in combination with frequency compounding].Ultraschall Med. 2007 Aug;28(4):387-93. doi: 10.1055/s-2007-963041. Epub 2007 Jun 29. Ultraschall Med. 2007. PMID: 17602368 German.
-
[Imaging of the abdominal wall: combination of tissue-harmonic-imaging and extended field-of-view sonography].Rofo. 2004 Dec;176(12):1837-42. doi: 10.1055/s-2004-813472. Rofo. 2004. PMID: 15573297 German.
-
Evaluation of enhancement patterns of focal nodular hyperplasia in contrast-enhanced, wide-band phase-inversion harmonic power Doppler imaging of the liver.Med Sci Monit. 2004 Jun;10 Suppl 3:17-21. Med Sci Monit. 2004. PMID: 16538193
-
Ultrasound in head and neck surgery: thyroid, parathyroid, and cervical lymph nodes.Surg Clin North Am. 2004 Aug;84(4):973-1000, v. doi: 10.1016/j.suc.2004.04.007. Surg Clin North Am. 2004. PMID: 15261750 Review.
-
A Primer on the Physical Principles of Tissue Harmonic Imaging.Radiographics. 2015 Nov-Dec;35(7):1955-64. doi: 10.1148/rg.2015140338. Radiographics. 2015. PMID: 26562232 Review.
Cited by
-
Prospective Comparison of Handheld Ultrasound Devices from Different Manufacturers with Respect to B-Scan Quality and Clinical Significance for Various Abdominal Sonography Questions.Diagnostics (Basel). 2023 Dec 8;13(24):3622. doi: 10.3390/diagnostics13243622. Diagnostics (Basel). 2023. PMID: 38132206 Free PMC article.
-
[Modern gray-scale sonography of the breast].Radiologe. 2010 Nov;50(11):964, 966-72. doi: 10.1007/s00117-010-2007-8. Radiologe. 2010. PMID: 20890590 Review. German.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Research Materials
Miscellaneous