Lithium dilution measurement of cardiac output and arterial pulse waveform analysis: an indicator dilution calibrated beat-by-beat system for continuous estimation of cardiac output
- PMID: 12386506
- DOI: 10.1097/00075198-200206000-00010
Lithium dilution measurement of cardiac output and arterial pulse waveform analysis: an indicator dilution calibrated beat-by-beat system for continuous estimation of cardiac output
Abstract
Lithium dilution cardiac output (LiDCO trade mark; LiDCO, London, UK) is a minimally invasive indicator dilution technique for the measurement of cardiac output. It was primarily developed as a simple calibration for the PulseCO trade mark (LiDCO, London, UK) continuous arterial waveform analysis monitor. The technique is quick and simple, requiring only an arterial line and central or peripheral venous access. These lines would probably already have been inserted in critical care patients. A small dose of lithium chloride is injected as an intravenous bolus, and cardiac output is derived from the dilution curve generated by a lithium-sensitive electrode attached to the arterial line. Studies in humans and animals have shown good agreement compared with results obtained with other techniques, and the efficacy of LiDCO trade mark in pediatric patients has also been proven. Compared with thermodilution, lithium dilution showed closer agreement in clinical studies with electromagnetic flow measurement.PulseCO trade mark is a beat-to-beat cardiac output monitor that calculates stroke volume from the arterial pressure waveform using an autocorrelation algorithm. The algorithm is not dependent on waveform morphology, but, rather, it calculates nominal stroke volume from a pressure-volume transform of the entire waveform. The nominal stroke volume is converted to actual stroke volume by calibration of the algorithm with LiDCO trade mark. Initial studies indicate good fidelity, and the results from centers in the United States and the United Kingdom are extremely encouraging. The PulseCO trade mark monitor incorporates software for interpretation of the hemodynamic data generated and provides a real-time analysis of arterial pressure variations (ie, stroke volume variation, pulse pressure variation, and systolic pressure variation) as theoretical guides to intravascular and cardiac filling.
Similar articles
-
Lithium dilution cardiac output measurement in the critically ill patient: determination of precision of the technique.Intensive Care Med. 2009 Mar;35(3):498-504. doi: 10.1007/s00134-008-1292-4. Epub 2008 Sep 18. Intensive Care Med. 2009. PMID: 18802681
-
Lithium dilution, pulse power analysis, and continuous thermodilution cardiac output measurements compared with bolus thermodilution in anaesthetized ponies.Br J Anaesth. 2012 Dec;109(6):864-9. doi: 10.1093/bja/aes269. Epub 2012 Aug 27. Br J Anaesth. 2012. PMID: 22926200
-
Comparison of continuous arterial pressure waveform analysis with the lithium dilution technique to monitor cardiac output in conscious dogs with systemic inflammatory response syndrome.Am J Vet Res. 2009 Nov;70(11):1365-73. doi: 10.2460/ajvr.70.11.1365. Am J Vet Res. 2009. PMID: 19878019
-
Equipment review: an appraisal of the LiDCO plus method of measuring cardiac output.Crit Care. 2004 Jun;8(3):190-5. doi: 10.1186/cc2852. Epub 2004 May 5. Crit Care. 2004. PMID: 15153237 Free PMC article. Review.
-
[Hemodynamic monitoring with the lithium dilution cardiac output system].Ugeskr Laeger. 2008 Feb 11;170(7):522-4. Ugeskr Laeger. 2008. PMID: 18291079 Review. Danish.
Cited by
-
Hemodynamic monitoring in the critically ill: an overview of current cardiac output monitoring methods.F1000Res. 2016 Dec 16;5:F1000 Faculty Rev-2855. doi: 10.12688/f1000research.8991.1. eCollection 2016. F1000Res. 2016. PMID: 28003877 Free PMC article. Review.
-
Bench-to-bedside review: Resuscitation in the emergency department.Crit Care. 2005 Apr;9(2):170-6. doi: 10.1186/cc2986. Epub 2004 Oct 20. Crit Care. 2005. PMID: 15774074 Free PMC article. Review.
-
Methods in pharmacology: measurement of cardiac output.Br J Clin Pharmacol. 2011 Mar;71(3):316-30. doi: 10.1111/j.1365-2125.2010.03798.x. Br J Clin Pharmacol. 2011. PMID: 21284692 Free PMC article. Review.
-
Real-time stroke volume measurements for the optimization of cardiac resynchronization therapy parameters.Europace. 2010 Sep;12(9):1270-4. doi: 10.1093/europace/euq175. Epub 2010 Jun 4. Europace. 2010. PMID: 20525728 Free PMC article.
-
Liberal or restricted fluid administration: are we ready for a proposal of a restricted intraoperative approach?BMC Anesthesiol. 2014 Aug 1;14:62. doi: 10.1186/1471-2253-14-62. eCollection 2014. BMC Anesthesiol. 2014. PMID: 25104915 Free PMC article. Clinical Trial.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical