Lithium dilution cardiac output measurement: a clinical assessment of central venous and peripheral venous indicator injection
- PMID: 12394944
- DOI: 10.1097/00003246-200210000-00004
Lithium dilution cardiac output measurement: a clinical assessment of central venous and peripheral venous indicator injection
Abstract
Objective: The lithium indicator dilution technique has been shown to measure cardiac output (CO) accurately by using central venous injection of lithium chloride (Li-CCO). This study aimed to compare the measurement of CO by using peripheral venous administration of lithium chloride (Li-PCO) with Li-CCO.
Design: Prospective, observational human study.
Setting: Surgical intensive care unit.
Patients: Thirty-one patients were studied after major surgery. All patients had arterial, central, and peripheral venous catheters. A total of 24 patients had pulmonary artery catheters.
Measurements: Serial measurements of Li-CCO and Li-PCO were made during hemodynamically stable conditions. CO was also measured using thermodilution (TDCO) when a pulmonary artery catheter was present. Data were analyzed by linear regression, the generalized estimating equation, and the comparison method described by Bland and Altman.
Main results: There were 93 Li-CCOs, 93 Li-PCOs, and 216 TDCOs recorded. The ranges of COs were similar: Li-CCO, 2.36-11.52 L/min (mean, 5.22 L/min; n = 31); Li-PCO, 1.63-9.99 L/min (mean, 5.22 L/min; n = 31), and TDCO, 3.28-10.4 L/min (mean, 5.75 L/min; n = 24). There was good linear correlation between Li-CCO and Li-PCO (R2 =.845). The mean difference for Li-CCO-Li-PCO was very small and insignificant (p =.97), and the limits of agreement were acceptable (mean difference +/- sd, 0.0005 +/- 0.64 L/min). The mean difference for Li-CCO-Li-PCO was smaller if the peripheral injection site was proximal rather than distal to the wrist (p =.053). Li-PCO and Li-CCO values were lower than simultaneously obtained TDCO measurements (Li-PCO-TDCO, -0.538 +/- 0.95 L/min, p =.003; Li-CCO-TDCO, -0.526 +/- 0.67 L/min, p =.0001).
Conclusions: Li-PCO gives a measurement that agrees well with Li-CCO. Accuracy of Li-PCO is probably improved if a proximal arm vein is used. Li-PCO provides accurate measurements of CO without the risks of pulmonary artery or central venous catheterization.
Comment in
-
Cardiac output measurement: is least invasive always the best?Crit Care Med. 2002 Oct;30(10):2380-2. doi: 10.1097/00003246-200210000-00035. Crit Care Med. 2002. PMID: 12394975 No abstract available.