PulseCO: a less-invasive method to monitor cardiac output from arterial pressure after cardiac surgery
- PMID: 12400827
- DOI: 10.1016/s0003-4975(02)04059-6
PulseCO: a less-invasive method to monitor cardiac output from arterial pressure after cardiac surgery
Abstract
Background: Cardiac output is often monitored after cardiac operations with a pulmonary artery catheter. A new method has been introduced that measures cardiac output by lithium dilution (LiDCO) and uses these data to calibrate a system (PulseCO) that calculates cardiac output continuously from the energy of the arterial pressure waveform. It is unknown whether PulseCO measurements are valid early after cardiac surgery when changes in temperature and vascular tone or intermittent use of the arterial line for blood sampling may occur. This study assessed the reliability of cardiac output determinations by PulseCO in the first 8 hours after cardiac surgery.
Methods: After a one-time PulseCO calibration, cardiac output was measured in 20 patients who had undergone coronary artery bypass grafting at 0, 2, 4, 6, and 8 hours after arrival in the intensive care unit using (1) thermodilution through a pulmonary artery catheter (Thermo); (2) lithium dilution (LiDCO); and (3) PulseCO. Concordance correlations were calculated between methods, and differences were compared by Wilcoxon paired rank test and Bland-Altman analysis.
Results: Cardiac output ranged from 3.4 to 8.5 L/min. No significant differences were noted between measurements obtained by each technique at any time point. Concordance correlations and Bland-Altman analysis confirmed good agreement between PulseCO and Thermo determinations of cardiac output during the study interval.
Conclusions: PulseCO measurements remain reliable without recalibration for at least 8 hours after cardiac surgery and may offer a less-invasive approach for early postoperative cardiac output monitoring.
Similar articles
-
Comparison of continuous arterial pressure waveform analysis with the lithium dilution technique to monitor cardiac output in conscious dogs with systemic inflammatory response syndrome.Am J Vet Res. 2009 Nov;70(11):1365-73. doi: 10.2460/ajvr.70.11.1365. Am J Vet Res. 2009. PMID: 19878019
-
Continuous cardiac output monitoring via arterial pressure waveform analysis following severe hemorrhagic shock in dogs.Crit Care Med. 2007 Jul;35(7):1724-9. doi: 10.1097/01.CCM.0000266590.25109.F2. Crit Care Med. 2007. PMID: 17452933
-
Lithium dilution measurement of cardiac output and arterial pulse waveform analysis: an indicator dilution calibrated beat-by-beat system for continuous estimation of cardiac output.Curr Opin Crit Care. 2002 Jun;8(3):257-61. doi: 10.1097/00075198-200206000-00010. Curr Opin Crit Care. 2002. PMID: 12386506 Review.
-
Comparison of thermodilution, lithium dilution, and pulse contour analysis for the measurement of cardiac output in 3 different hemodynamic states in dogs.J Vet Emerg Crit Care (San Antonio). 2014 Sep-Oct;24(5):562-70. doi: 10.1111/vec.12219. Epub 2014 Aug 20. J Vet Emerg Crit Care (San Antonio). 2014. PMID: 25142925
-
Pressure Waveform Analysis.Anesth Analg. 2018 Jun;126(6):1930-1933. doi: 10.1213/ANE.0000000000002527. Anesth Analg. 2018. PMID: 29077613 Review.
Cited by
-
A Prospective Study to Evaluate the Accuracy of Pulse Power Analysis to Monitor Cardiac Output in Critically Ill Patients.BMC Anesthesiol. 2008 Feb 18;8:3. doi: 10.1186/1471-2253-8-3. BMC Anesthesiol. 2008. PMID: 18282275 Free PMC article.
-
Reliability of continuous cardiac output measurement during intra-abdominal hypertension relies on repeated calibrations: an experimental animal study.Crit Care. 2008;12(5):R132. doi: 10.1186/cc7102. Epub 2008 Oct 29. Crit Care. 2008. PMID: 18957114 Free PMC article.
-
Cross-comparison of cardiac output trending accuracy of LiDCO, PiCCO, FloTrac and pulmonary artery catheters.Crit Care. 2010;14(6):R212. doi: 10.1186/cc9335. Epub 2010 Nov 23. Crit Care. 2010. PMID: 21092290 Free PMC article.
-
Estimation of cardiac output in a pharmacological trial using a simple method based on arterial blood pressure signal waveform: a comparison with pulmonary thermodilution and echocardiographic methods.Eur J Clin Pharmacol. 2006 Jun;62(6):401-7. doi: 10.1007/s00228-006-0115-1. Epub 2006 Mar 29. Eur J Clin Pharmacol. 2006. PMID: 16570186 Clinical Trial.
-
Non-Invasive Device for Blood Pressure Wave Acquisition by Means of Mechanical Transducer.Sensors (Basel). 2019 Oct 5;19(19):4311. doi: 10.3390/s19194311. Sensors (Basel). 2019. PMID: 31590351 Free PMC article.