Case Reports
. 2002 Nov;48(11):2023-9.

Wrong biochemistry results: two case reports and observational study in 5310 patients on potentially misleading thyroid-stimulating hormone and gonadotropin immunoassay results

Affiliations
  • PMID: 12406989
Case Reports

Wrong biochemistry results: two case reports and observational study in 5310 patients on potentially misleading thyroid-stimulating hormone and gonadotropin immunoassay results

Adel A A Ismail et al. Clin Chem. 2002 Nov.

Abstract

Background: Immunoassays are used in almost all medical and surgical specialties, but they suffer from interference from proteins such as antibodies in some patients' sera. Such interferences are usually reported in the literature only as case reports after the introduction of a new assay.

Methods: We undertook a prospective observational study on 5310 patients for whom the common immunoassay tests for thyroid-stimulating hormone (TSH) and/or gonadotropins were requested. All TSH and gonadotropin results were critically assessed for a mismatch between the clinical details and analytical results to identify samples suspected of analytical unreliability. These were tested further by three approaches to screen for interference.

Results: From the 5310 sets of results, 59 patients' samples were identified as suspect and were tested further. Analytically incorrect results were found in 28 (0.53% of the total studied). The magnitude of interference varied, but in 23 of 28 patients (82%), it was considered large enough to have a potentially adverse effect on cost and/or the clinical care of these patients. Two cases, described in detail, illustrate the adverse effect of error on patient care and cost, and the second highlights the difficulties and limitations of current approaches for identifying interference and inaccuracy in immunoassays.

Conclusions: Because millions of TSH/gonadotropin tests are carried out in UK hospital laboratories alone, our data suggest that thousands of patients could be adversely affected by errors from interferences. Early identification of interference in cases with unusual results could be valuable.

PubMed Disclaimer

Comment in

  • Striving for perfection.
    Leape LL. Leape LL. Clin Chem. 2002 Nov;48(11):1871-2. Clin Chem. 2002. PMID: 12406970 No abstract available.

Publication types

LinkOut - more resources