Accuracy of displayed values of tidal volume in the pediatric intensive care unit
- PMID: 12441771
- DOI: 10.1097/00003246-200211000-00027
Accuracy of displayed values of tidal volume in the pediatric intensive care unit
Abstract
Objectives: To assess the accuracy of the expired tidal volumes (VT(E)) displayed by one of the most frequently used ventilators that measures exhaled volume at the expiratory valve.
Design: Prospective study.
Setting: The intensive care units of a pediatric tertiary referral center in London, UK.
Patients: A total of 56 intubated children aged between 3 wks and 16.6 yrs who were clinically stable and ventilated with a Servo 300 ventilator.
Interventions: The CO2SMO Plus respiratory monitor, which measures flow at the airway opening, was validated using calibrated syringes and appropriate tracheal tubes and connections. Simultaneous in vivo recordings of VT(E) from the Servo 300 and CO2SMO Plus were compared before (displayed Servo VT(E)) and after (effective Servo VT(E)) compensating for ventilator circuit compliance.
Measurements and main results: The in vitro accuracy of the CO2SMO Plus was within +/-5% over a wide range of volumes and measurement conditions. The displayed Servo 300 VT(E) overestimated the true VT(E) by between 2% and 91%. The magnitude of error varied within and between children, according to pressure change (peak inspiratory pressure minus positive end-expiratory pressure), VT(E), and circuit size. Mean (sd) error was 32% (20%) in 40 children with displayed Servo VT(E) of <160 mL and 18% (6%) in 16 subjects with displayed Servo VT(E) of >/=160 mL. After correcting for gas compression, effective VT(E) from the Servo 300 underestimated the true VT(E) by up to 64% in the smallest infants but continued to overestimate by as much as 29% in older children.
Conclusions: The accuracy of tidal volume values is crucially dependent on the site of measurement. Unless measured at the airway opening, displayed values are an inconsistent and misleading indicator of the true volumes delivered.
Comment in
-
Bedside tidal volume measurements: GIGO?Crit Care Med. 2002 Nov;30(11):2606. doi: 10.1097/00003246-200211000-00041. Crit Care Med. 2002. PMID: 12441785 No abstract available.
Similar articles
-
Tidal volumes for ventilated infants should be determined with a pneumotachometer placed at the endotracheal tube.Am J Respir Crit Care Med. 2000 Dec;162(6):2109-12. doi: 10.1164/ajrccm.162.6.9906112. Am J Respir Crit Care Med. 2000. PMID: 11112123
-
Reliability of Displayed Tidal Volume in Healthy and Surfactant-Depleted Piglets.Respir Care. 2016 Dec;61(12):1605-1612. doi: 10.4187/respcare.04569. Epub 2016 Nov 15. Respir Care. 2016. PMID: 27899539
-
Reliability of displayed tidal volume in infants and children during dual-controlled ventilation.Pediatr Crit Care Med. 2009 Nov;10(6):661-7. doi: 10.1097/PCC.0b013e3181bb2b2b. Pediatr Crit Care Med. 2009. PMID: 19851123
-
[What are the changes in paediatric anaesthesia practice afforded by new anaesthetic ventilators?].Ann Fr Anesth Reanim. 2006 Apr;25(4):417-23. doi: 10.1016/j.annfar.2005.10.003. Epub 2005 Nov 28. Ann Fr Anesth Reanim. 2006. PMID: 16310339 Review. French.
-
Assessment of lung function in the intensive care unit.Pediatr Pulmonol. 1995 Feb;19(2):118-28. doi: 10.1002/ppul.1950190207. Pediatr Pulmonol. 1995. PMID: 7659467 Review.
Cited by
-
Influence of compliance and resistance of the test lung on the accuracy of the tidal volume delivered by the ventilator.BMC Pulm Med. 2024 Oct 9;24(1):498. doi: 10.1186/s12890-024-03294-1. BMC Pulm Med. 2024. PMID: 39385191 Free PMC article.
-
Trends in mechanical ventilation: are we ventilating our patients in the best possible way?Breathe (Sheff). 2017 Jun;13(2):84-98. doi: 10.1183/20734735.007817. Breathe (Sheff). 2017. PMID: 28620428 Free PMC article. Review.
-
The influence of physiotherapy and suction on respiratory deadspace in ventilated children.Intensive Care Med. 2004 Jun;30(6):1152-9. doi: 10.1007/s00134-004-2261-1. Epub 2004 May 8. Intensive Care Med. 2004. PMID: 15138674 Clinical Trial.
-
Is volume and leak monitoring feasible during nasopharyngeal continuous positive airway pressure in neonates?Intensive Care Med. 2009 Nov;35(11):1934-41. doi: 10.1007/s00134-009-1651-9. Intensive Care Med. 2009. PMID: 19771409 Clinical Trial.
-
Comparison of volumetric capnography and mixed expired gas methods to calculate physiological dead space in mechanically ventilated ICU patients.Intensive Care Med. 2012 Oct;38(10):1712-7. doi: 10.1007/s00134-012-2670-5. Epub 2012 Aug 15. Intensive Care Med. 2012. PMID: 22893221
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources