Filtration characteristics of dacron wool (Swank) blood transfusion filters
- PMID: 124463
- DOI: 10.1097/00007611-197506000-00008
Filtration characteristics of dacron wool (Swank) blood transfusion filters
Abstract
Stored human blood of varying age was passed through standard commercial blood transfusion filters (pore size, 170 mu) and Swank Dacron wool blood transfusion filters (pore size, 20mu). Passage through the Dacron filter resulted in a marked decrease in SFP and an increase in filter weight indicating removal of microaggregates which have been implicated as a cause of pulmonary insufficiency. The commercial filter tested did not appear to be effective in removing these harmful aggregates. On the basis of this research it is concluded that the Swank Dacron wool blood transfusion filter will prevent pulmonary microemboli during transfusion. Because aggregate removal appeared to be quantitative, it is recommended that no more than four units of blood be passed through each Dacron wool filter.
Similar articles
-
A comparison in vivo dacron wool (Swank) and polyester mesh (Pall) micropore blood transfusion filters in the prevention of pulmonary microembolism associated with massive transfusion.Ann Surg. 1975 Dec;182(6):690-5. doi: 10.1097/00000658-197512000-00005. Ann Surg. 1975. PMID: 242282 Free PMC article.
-
Filtration characteristics of polyester mesh (Pall) micropore blood transfusion filter.Arch Surg. 1976 Jan;111(1):56-9. doi: 10.1001/archsurg.1976.01360190058009. Arch Surg. 1976. PMID: 128340
-
Increased pulmonary arteriovenous shunting in humans following blood transfusion. Relation to screen filtration pressure of transfused blood and prevention by Dacron wool (Swank) filtration.Arch Surg. 1978 Aug;113(8):947-50. doi: 10.1001/archsurg.1978.01370200041008. Arch Surg. 1978. PMID: 150833
-
The filtration of blood.Surg Annu. 1978;10:105-22. Surg Annu. 1978. PMID: 147523 Review. No abstract available.
-
[Origin and significance of particles causing microemboli and clinical effectiveness of using filters in artificial blood circulation equipment].Med Tekh. 1985 Sep-Oct;(5):57-61. Med Tekh. 1985. PMID: 3908878 Review. Russian. No abstract available.
Cited by
-
A comparison in vivo dacron wool (Swank) and polyester mesh (Pall) micropore blood transfusion filters in the prevention of pulmonary microembolism associated with massive transfusion.Ann Surg. 1975 Dec;182(6):690-5. doi: 10.1097/00000658-197512000-00005. Ann Surg. 1975. PMID: 242282 Free PMC article.
-
Progression and resolution of changes in pulmonary function and structure due to pulmonary microembolism and blood transfusion.Ann Surg. 1977 Jan;185(1):92-99. doi: 10.1097/00000658-197701000-00016. Ann Surg. 1977. PMID: 831640 Free PMC article.
-
Microaggregate formation in stored human packed cells: comparison with formation in stored whole blood and a method for their removal.Ann Surg. 1976 Feb;183(2):109-13. doi: 10.1097/00000658-197602000-00004. Ann Surg. 1976. PMID: 1247308 Free PMC article.