A close-up look at an implant fracture: a case report
- PMID: 12449308
A close-up look at an implant fracture: a case report
Abstract
The authors report on the fracture of a standard endosseous dental implant 11 years after placement. It had been connected to a natural tooth using a nonrigid connector to compensate for the lack of a built-in antirotational system on the prosthetic abutment of the implant restoration. Scanning electron microscopy revealed striations on the fracture surface, suggesting a fatigue-associated failure.
Similar articles
-
Dental implants as strategic supplementary abutments for implant-tooth-supported telescopic crown-retained maxillary dentures: a retrospective follow-up study for up to 9 years.Int J Prosthodont. 2007 Nov-Dec;20(6):617-22. Int J Prosthodont. 2007. PMID: 18069371
-
Scanning electron microscopy observations of failures of implant overdenture bars: a case series report.Clin Implant Dent Relat Res. 2010 Mar;12(1):26-38. doi: 10.1111/j.1708-8208.2008.00127.x. Clin Implant Dent Relat Res. 2010. PMID: 20148915
-
Retrospective evaluation of complete-arch fixed partial dentures connecting teeth and implant abutments in patients with normal and reduced periodontal support.J Prosthet Dent. 2005 Oct;94(4):313-20. doi: 10.1016/j.prosdent.2005.08.007. J Prosthet Dent. 2005. PMID: 16198167
-
Intrusion phenomenon in combination tooth-implant restorations: a review of the literature.J Prosthet Dent. 1998 Aug;80(2):199-203. doi: 10.1016/s0022-3913(98)70110-6. J Prosthet Dent. 1998. PMID: 9710822 Review.
-
Management of Implant/Prosthodontic Complications.Dent Clin North Am. 2019 Apr;63(2):217-231. doi: 10.1016/j.cden.2018.11.004. Epub 2019 Feb 5. Dent Clin North Am. 2019. PMID: 30825987 Review.
Cited by
-
Dental implant fractures - aetiology, treatment and case report.J Clin Diagn Res. 2014 Mar;8(3):300-4. doi: 10.7860/JCDR/2014/8074.4158. Epub 2014 Mar 15. J Clin Diagn Res. 2014. PMID: 24783165 Free PMC article. Review.
-
Effect of loading frequency on cyclic fatigue lifetime of a standard-diameter implant with an internal abutment connection.Dent Mater. 2018 Dec;34(12):1711-1716. doi: 10.1016/j.dental.2018.09.001. Epub 2018 Sep 13. Dent Mater. 2018. PMID: 30220505 Free PMC article.