. 2002 Dec;24(6):757-67.
doi: 10.1097/00007691-200212000-00013.

Simultaneous and rapid analysis of cyclosporin A and creatinine in finger prick blood samples using liquid chromatography tandem mass spectrometry and its application in C2 monitoring

Affiliations

Simultaneous and rapid analysis of cyclosporin A and creatinine in finger prick blood samples using liquid chromatography tandem mass spectrometry and its application in C2 monitoring

Brian G Keevil et al. Ther Drug Monit. 2002 Dec.

Abstract

A simple and rapid liquid chromatography tandem mass spectrometry (LC-MS/MS) method for the simultaneous analysis of cyclosporin A (CsA) and creatinine using capillary blood has been developed. Venous and capillary blood samples were taken predose and at C2 from 65 heart and lung transplant recipients (65 x 4 samples). For comparisons, serum creatinine and blood CsA concentrations were measured by the Jaffe and EMIT methods, respectively, using an Olympus AU600 analyzer. For the LC-MS/MS assay, samples were prepared in a 96 x 700-microL well block by adding 10 microL of blood (or serum) to 40 microL of 0.1 mol/L zinc sulphate solution containing deuterated creatinine internal standard. Proteins were precipitated by adding 100 microL acetonitrile containing ascomycin internal standard. After vigorous mixing and centrifugation, 5 microL of the supernatant was injected into the LC-MS/MS system. A Waters 2795 high-performance liquid chromatography (HPLC) system was used to elute a C18 cartridge (3 mm x 4 mm) at 0.6 mL/min with a step gradient of 50-100% methanol containing 2 mmol/L ammonium acetate and 0.1% (v/v) formic acid. The column was maintained at 55 degrees C, and the retention times were creatinine, 0.4 minutes; ascomycin, 0.98 minutes; and CsA, 1.2 minutes. Cycle time was 2.5 minutes, injection to injection. The analytes were monitored using a Quattro microtandem mass spectrometer operated in multiple reaction monitoring mode using the following transitions: creatinine, m/z 114>44; d3-creatinine (IS), m/z 117>47; ascomycin (IS), m/z 809>756; and CsA, m/z 1,220>1,203. Assay characteristics were CsA intraassay CV, 3.6-3.0% (33-1,500 microg/L); CsA interassay CV, 6.7-2.5% (10-5,000 microg/L); LC-MS/MS capillary [CsA] = 0.99 x LC-MS/MS venous [CsA] - 4.2, R = 0.98; and LC-MS/MS venous [CsA] = 0.93 x EMIT venous [CsA] + 2.9, R = 0.98. Creatinine intraassay CV, 6.6-2.5% (20-720 micromol/L); interassay CV, 5.7-3.3% (80-590 micromol/L); LC-MS/MS capillary [creatinine] = 0.99 Jaffe plasma [creatinine] -42.6, R = 0.87. Total time for the preparation and analysis of 30 samples was approximately 2 hours. This assay will provide a flexible, robust, and cost-effective solution for the monitoring of CsA and creatinine in transplant recipients with potential applications in pediatric medicine and pharmacokinetic studies, in which frequent sampling is required.

PubMed Disclaimer

Similar articles

Cited by

LinkOut - more resources