Biomedical implants and devices: assessment of magnetic field interactions with a 3.0-Tesla MR system

Affiliations
Comparative Study

Biomedical implants and devices: assessment of magnetic field interactions with a 3.0-Tesla MR system

Frank G Shellock. J Magn Reson Imaging. 2002 Dec.

Abstract

Purpose: To evaluate magnetic field interactions for 109 different biomedical implants and devices in association with exposure to a 3.0-Tesla magnetic resonance (MR) system.

Materials and methods: A total of 109 implants and devices (aneurysm clips, 32; clips, fasteners, and staples, 10; coils and stents, 10; heart valve prostheses and annuloplasty rings, 12; orthopedic implants, five; suture materials, 13; vascular access ports and accessories, 13; miscellaneous implants and devices, 14) were tested for magnetic field interactions at 3.0-Tesla using previously-described, standardized techniques to assess magnetic field translational attraction and torque.

Results: The deflection angles and torque measurements ranged, respectively, from 0 to 16 degrees and 0 to +2 for the aneurysm clips; 0 to 90 degrees and 0 to +4 for the clips, fasteners, and staples; 0 to 47 degrees and 0 to +4 for the coils and stents; 0 to 4 degrees and 0 to +1 for the heart valve prostheses and annuloplasty rings; 0 to 12 degrees and 0 to +2 for the orthopedic implants; 0 to 13 degrees and 0 to +2 for the suture materials; 0 to 52 degrees and 0 to +4 for the vascular access ports and accessories; and 0 to 28 degrees and 0 to +3 for the miscellaneous implants and devices.

Conclusion: Of the 109 implants and devices assessed for magnetic field interactions at 3.0-Tesla, four (4%) are potentially unsafe based on deflection angle criteria. The implications of these results for patients undergoing MR procedures at 3.0-Tesla is discussed. Notably, these results are specific to the 3.0-Tesla MR system used for this evaluation.