Real-time pulse oximetry artifact annotation on computerized anaesthetic records
- PMID: 12455744
- DOI: 10.1023/a:1020785323889
Real-time pulse oximetry artifact annotation on computerized anaesthetic records
Abstract
Objectives: Adoption of computerised anaesthesia record keeping systems has been limited by the concern that they record artifactual data and accurate data indiscriminately. Data resulting from artifacts does not reflect the patient's true condition and presents a problem in later analysis of the record, with associated medico-legal implications. This study developed an algorithm to automatically annotate pulse oximetry artifacts and sought to evaluate the algorithm's accuracy in routine surgical procedures.
Methods: MacAnaesthetist is a semi-automatic anaesthetic record keeping system developed for the Apple Macintosh computer, which incorporated an algorithm designed to automatically detect pulse oximetry artifacts. The algorithm labeled artifactual oxygen saturation values < 90%. This was done in real-time by analyzing physiological data captured from a Datex AS/3 Anaesthesia Monitor. An observational study was conducted to evaluate the accuracy of the algorithm during routine surgical procedures (n = 20). An anaesthetic record was made by an anaesthetist using the Datex AS/3 record keeper, while a second anaesthetic record was produced in parallel using MacAnaesthetist. A copy of the Datex AS/3 record was kept for later review by a group of anaesthetists (n = 20), who judged oxygen saturation values < 90% to be either genuine or artifact.
Results: MacAnaesthetist correctly labeled 12 out of 13 oxygen saturations < 90% (92.3% accuracy). A post-operative review of the Datex AS/3 anaesthetic records (n = 8) by twenty anaesthetists resulted in 127 correct responses out of total of 200 (63.5% accuracy). The remaining Datex AS/3 records (n = 12) were not reviewed, as they did not contain any oxygen saturations <90%.
Conclusions: The real-time artifact detection algorithm developed in this study was more accurate than anaesthetists who post-operatively reviewed records produced by an existing computerised anaesthesia record keeping system. Algorithms have the potential to more accurately identify and annotate artifacts on computerised anaesthetic records, assisting clinicians to more correctly interpret abnormal data.
Similar articles
-
Does manual anaesthetic record capture remove clinically important data?Br J Anaesth. 2011 Oct;107(4):546-52. doi: 10.1093/bja/aer163. Epub 2011 Jun 10. Br J Anaesth. 2011. PMID: 21665898
-
[Online recording of monitor data. The artifact problem].Anaesthesist. 1995 Dec;44(12):818-25. doi: 10.1007/s001010050217. Anaesthesist. 1995. PMID: 8594955 German.
-
Two methods to adapt the human haemoglobin-oxygen dissociation algorithm to the blood of white rhinoceros (Ceratotherium simum) and to determine the accuracy of pulse oximetry.Vet Anaesth Analg. 2016 Sep;43(5):566-70. doi: 10.1111/vaa.12334. Epub 2016 Jan 29. Vet Anaesth Analg. 2016. PMID: 26821767
-
Patient data management systems in anaesthesia: an emerging technology.Can J Anaesth. 1995 Oct;42(10):914-21. doi: 10.1007/BF03011040. Can J Anaesth. 1995. PMID: 8706202 Review.
-
The problem of artifacts in patient monitor data during surgery: a clinical and methodological review.Anesth Analg. 2006 Nov;103(5):1196-204. doi: 10.1213/01.ane.0000247964.47706.5d. Anesth Analg. 2006. PMID: 17056954 Review.
Cited by
-
Artifacts in research data obtained from an anesthesia information and management system.Can J Anaesth. 2012 Sep;59(9):833-41. doi: 10.1007/s12630-012-9754-0. Epub 2012 Jul 18. Can J Anaesth. 2012. PMID: 22806063 Free PMC article.
-
Automated anesthesia artifact analysis: can machines be trained to take out the garbage?J Clin Monit Comput. 2021 Apr;35(2):225-227. doi: 10.1007/s10877-020-00589-6. Epub 2020 Sep 12. J Clin Monit Comput. 2021. PMID: 32920701 No abstract available.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials