A comparison of patient state index and bispectral index values during the perioperative period
- PMID: 12456436
- DOI: 10.1097/00000539-200212000-00036
A comparison of patient state index and bispectral index values during the perioperative period
Abstract
The patient state index (PSI), a quantitative electroencephalographic index, has been recently introduced into clinical practice as a monitor for assessing consciousness during sedation and general anesthesia. We designed this observational study to compare the sensitivity and specificity of the PSI with that of the bispectral index (BIS) with respect to their ability to predict the loss of consciousness and emergence from anesthesia, as well as to assess changes in IV (propofol) and inhaled (desflurane) anesthetics during the maintenance period. Twenty consenting patients scheduled for elective laparoscopic surgical procedures were enrolled in this prospective clinical study. Anesthesia was induced with propofol 2 mg/kg IV and fentanyl 1 micro g/kg IV, and tracheal intubation was facilitated with cisatracurium 0.3 mg/kg IV. Desflurane 4% in combination with nitrous oxide 60% in oxygen was administered for the maintenance of anesthesia. Comparative PSI and BIS values were obtained at specific time intervals during the induction, maintenance, and emergence periods. The changes in these indices were recorded after the administration of propofol (20 mg IV) or with 2% increases or decreases in the inspired concentration of desflurane during the maintenance period. With logistic regression models, both the BIS and PSI were found to be effective as predictors of unconsciousness (i.e., failed to respond to verbal stimuli) (P < 0.01). The PSI also correlated with the BIS during both the induction of (r = 0.78) and emergence from (r = 0.73) general anesthesia. However, the area under the receiver operating characteristic curve for detection of consciousness indicated a better performance with the PSI (0.95 +/- 0.04) than the BIS (0.79 +/- 0.04). During the maintenance period, the PSI values were comparable to the BIS in response to changes in propofol and desflurane but displayed greater interpatient variability. Finally, the PSI (versus BIS) values were less interfered with by the electrocautery unit during surgery (16% versus 65%, respectively). In conclusion, the PSI may prove to be a viable alternative to the BIS for evaluating consciousness during the induction of and emergence from general anesthesia, as well as for titrating the administration of propofol and desflurane during the maintenance period. However, further studies with the PSA device are needed to determine its role in anesthesia.
Implications: The patient state index could be a useful alternative to the bispectral index for assessing level of consciousness during the induction of and emergence from anesthesia, as well as for titrating IV and volatile anesthetics during surgery.
Similar articles
-
Is the patient state analyzer with the PSArray2 a cost-effective alternative to the bispectral index monitor during the perioperative period?Anesth Analg. 2004 Nov;99(5):1429-1435. doi: 10.1213/01.ANE.0000132784.57622.CC. Anesth Analg. 2004. PMID: 15502043
-
A comparison of state and response entropy versus bispectral index values during the perioperative period.Anesth Analg. 2006 Jan;102(1):160-7. doi: 10.1213/01.ane.0000183668.53139.fc. Anesth Analg. 2006. PMID: 16368823
-
Is the bispectral index useful in predicting fast-track eligibility after ambulatory anesthesia with propofol and desflurane?Anesth Analg. 1998 Dec;87(6):1245-8. doi: 10.1097/00000539-199812000-00006. Anesth Analg. 1998. PMID: 9842806 Clinical Trial.
-
Does the use of electroencephalographic bispectral index or auditory evoked potential index monitoring facilitate recovery after desflurane anesthesia in the ambulatory setting?Anesthesiology. 2004 Apr;100(4):811-7. doi: 10.1097/00000542-200404000-00010. Anesthesiology. 2004. PMID: 15087615 Clinical Trial.
-
Towards a better understanding of anesthesia emergence mechanisms: Research and clinical implications.World J Methodol. 2018 Oct 12;8(2):9-16. doi: 10.5662/wjm.v8.i2.9. eCollection 2018 Oct 12. World J Methodol. 2018. PMID: 30345225 Free PMC article. Review.
Cited by
-
Rapid arousal from anaesthesia after reversal of deep rocuronium-induced neuromuscular block with sugammadex in a neuroradiological procedure.BMJ Case Rep. 2021 Aug 20;14(8):e242820. doi: 10.1136/bcr-2021-242820. BMJ Case Rep. 2021. PMID: 34417232 Free PMC article.
-
Comparison of bispectral index and patient state index as measures of sedation depth during surgeries using remimazolam tosilate.BMC Anesthesiol. 2023 Jun 15;23(1):208. doi: 10.1186/s12871-023-02172-3. BMC Anesthesiol. 2023. PMID: 37322424 Free PMC article.
-
Evaluation of the SEDline to improve the safety and efficiency of conscious sedation.Proc (Bayl Univ Med Cent). 2011 Jul;24(3):200-4. doi: 10.1080/08998280.2011.11928715. Proc (Bayl Univ Med Cent). 2011. PMID: 21738291 Free PMC article.
-
Awareness and recall during general anesthesia.Korean J Anesthesiol. 2014 May;66(5):339-45. doi: 10.4097/kjae.2014.66.5.339. Epub 2014 May 26. Korean J Anesthesiol. 2014. PMID: 24910724 Free PMC article. Review.
-
Evaluation of Sedation Levels Using SedLine During Intravenous Sedation for Dental Procedures: A Case-Series Study.Anesth Prog. 2023 Jun 1;70(2):85-87. doi: 10.2344/anpr-70-01-01. Anesth Prog. 2023. PMID: 37379089 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Research Materials
Miscellaneous