Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2002 Nov-Dec;18(6):683-91.
doi: 10.3928/1081-597X-20021101-04.

A new intraocular lens design to reduce spherical aberration of pseudophakic eyes

Affiliations
Comparative Study

A new intraocular lens design to reduce spherical aberration of pseudophakic eyes

Jack T Holladay et al. J Refract Surg. 2002 Nov-Dec.

Abstract

Purpose: The aim of this study was to design and evaluate in the laboratory a new intraocular lens (IOL) intended to provide superior ocular optical quality by reducing spherical aberration.

Methods: Corneal topography measurements were performed on 71 cataract patients using an Orbscan I. The measured corneal surface shapes were used to determine the wavefront aberration of each cornea. A model cornea was then designed to reproduce the measured average spherical aberration. This model cornea was used to design IOLs having a fixed amount of negative spherical aberration that partially compensates for the average positive spherical aberration of the cornea. Theoretical and physical eye models were used to assess the expected improvement in optical quality of an eye implanted with this lens.

Results: Measurements of optical quality provided evidence that if this modified prolate IOL was centered within 0.4 mm and tilted less than 7 degrees, it would exceed the optical performance of a conventional spherical IOL. This improvement occurred without an apparent loss in depth of focus.

Conclusion: A new IOL with a prolate anterior surface, designed to partially compensate for the average spherical aberration of the cornea, is intended to improve the ocular optical quality of pseudophakic patients.

PubMed Disclaimer