The influence of glenohumeral prosthetic mismatch on glenoid radiolucent lines: results of a multicenter study
- PMID: 12473707
- DOI: 10.2106/00004623-200212000-00010
The influence of glenohumeral prosthetic mismatch on glenoid radiolucent lines: results of a multicenter study
Abstract
Background: In shoulder arthroplasty, mismatch is defined as the difference in the radius or diameter of curvature between the humeral head and glenoid components. Recommendations for mismatch have not been substantiated scientifically. The purpose of this study was to evaluate the effect of mismatch on glenoid radiolucent lines.
Methods: The results of 319 total shoulder arthroplasties performed for the treatment of primary osteoarthritis were evaluated. All of the arthroplasties were performed with a single type of prosthesis (Aequalis; Tornier, Montbonnot, France) that included a cemented, all-polyethylene glenoid component. Three sizes of glenoid components and seven humeral head diameters were utilized. Radial mismatch was categorized as < or = 4 mm, 4.5 to 5.5 mm, 6 to 7 mm, or >7 to 10 mm. Radiographs were evaluated at a mean of 53.5 months (range, twenty-four to 110 months) postoperatively. Glenoid radiolucent lines were scored with a scale ranging from 0 points for no radiolucency to 18 points for radiolucent lines exceeding 2 mm in six zones. Variance, linear contrasts polynomial, quadratic polynomial contrast statistical, and linear regression analyses were performed to evaluate the relationship between radial mismatch and glenoid radiolucent lines.
Results: A significant linear relationship was found between mismatch and the glenoid radiolucency score (p < 0.0001), with significantly lower (better) radiolucency scores associated with radial mismatches of >5.5 mm.
Conclusions: In this study of glenohumeral prosthetic mismatch ranging from 0 to 10 mm, the mismatch had a significant influence on the scores for the glenoid radiolucent lines, which were best when the radial mismatch was between 6 and 10 mm. The theoretical risk of prosthetic instability with larger mismatch values was not demonstrated within the range of mismatch values evaluated in this series.
Similar articles
-
Patterns of loosening of polyethylene keeled glenoid components after shoulder arthroplasty for primary osteoarthritis: results of a multicenter study with more than five years of follow-up.J Bone Joint Surg Am. 2012 Jan 18;94(2):145-50. doi: 10.2106/JBJS.J.00699. J Bone Joint Surg Am. 2012. PMID: 22258001
-
The association of incomplete glenoid component seating and periprosthetic glenoid radiolucencies after total shoulder arthroplasty.J Shoulder Elbow Surg. 2016 Mar;25(3):442-7. doi: 10.1016/j.jse.2015.08.005. Epub 2015 Oct 9. J Shoulder Elbow Surg. 2016. PMID: 26456426
-
Reverse total shoulder arthroplasty for primary glenohumeral osteoarthritis in patients with a biconcave glenoid.J Bone Joint Surg Am. 2013 Jul 17;95(14):1297-304. doi: 10.2106/JBJS.L.00820. J Bone Joint Surg Am. 2013. PMID: 23864178
-
Glenoid morphology affects the incidence of radiolucent lines around cemented pegged polyethylene glenoid components.Arch Orthop Trauma Surg. 2013 Oct;133(10):1331-9. doi: 10.1007/s00402-013-1813-7. Epub 2013 Jul 14. Arch Orthop Trauma Surg. 2013. PMID: 23852591 Clinical Trial.
-
Total shoulder arthroplasty: glenoid component design.J Shoulder Elbow Surg. 2005 Jan-Feb;14(1 Suppl S):122S-128S. doi: 10.1016/j.jse.2004.09.028. J Shoulder Elbow Surg. 2005. PMID: 15726072 Review.
Cited by
-
New design of a cementless glenoid component in unconstrained shoulder arthroplasty: a prospective medium-term analysis of 143 cases.Eur J Orthop Surg Traumatol. 2013 Jan;23(1):27-34. doi: 10.1007/s00590-012-1109-6. Epub 2012 Oct 27. Eur J Orthop Surg Traumatol. 2013. PMID: 23293576 Free PMC article.
-
Assessment of radiolucent lines in cemented shoulder hemi-arthroplasties: study of concordance and reproducibility.Int Orthop. 2009 Feb;33(1):165-9. doi: 10.1007/s00264-007-0452-4. Epub 2007 Oct 17. Int Orthop. 2009. PMID: 17940769 Free PMC article.
-
The effect of radial mismatch on radiographic glenoid loosening.JSES Open Access. 2019 Nov 18;3(4):287-291. doi: 10.1016/j.jses.2019.09.007. eCollection 2019 Dec. JSES Open Access. 2019. PMID: 31891027 Free PMC article.
-
Radiographic performance depends on the radial glenohumeral mismatch in total shoulder arthroplasty.BMC Musculoskelet Disord. 2020 Apr 3;21(1):206. doi: 10.1186/s12891-020-03219-z. BMC Musculoskelet Disord. 2020. PMID: 32245455 Free PMC article.
-
Glenoid or not glenoid component in primary osteoarthritis.Eur J Orthop Surg Traumatol. 2013 May;23(4):387-93. doi: 10.1007/s00590-012-1117-6. Epub 2012 Nov 30. Eur J Orthop Surg Traumatol. 2013. PMID: 23412243 Review.