Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2002 Dec;17(8):987-91.
doi: 10.1054/arth.2002.35797.

A study of polyethylene and modularity issues in >1000 posterior cruciate-retaining knees at 5 to 11 years

Affiliations
Comparative Study

A study of polyethylene and modularity issues in >1000 posterior cruciate-retaining knees at 5 to 11 years

Andrew B Weber et al. J Arthroplasty. 2002 Dec.

Abstract

Polyethylene quality and production technique and monoblock versus modular design of tibial components affect the polyethylene wear of total knee arthroplasties (TKAs). We reviewed >1000 TKAs performed with the AGC prosthesis (Biomet, Warsaw, IN) using a uniform surgical technique with a 5- to 11-year follow-up. Of tibial components, 698 were of compression-molded monoblock design and 353 were of ram-extruded modular assembly design. The effect of the differences in tibial component design on the 5- to 11-year clinical and radiologic outcomes was studied. Higher rates of osteolysis, radiolucent lines, and revision were found with the ram-extruded modular design. This modular design is similar to most tibial component designs currently used in TKA, yet osteolysis in completely cemented components is described rarely. The TKAs done with a compression-molded monoblock design gave better clinical and radiologic results. The improved results may be due to polyethylene quality, assembly design, or a combination of these factors.

PubMed Disclaimer

Similar articles

Cited by

Publication types

LinkOut - more resources