Validation of the Microlife BP 3BTO-A oscillometric blood pressure monitoring device according to a modified British Hypertension Society protocol
- PMID: 12488652
- DOI: 10.1097/00126097-200212000-00005
Validation of the Microlife BP 3BTO-A oscillometric blood pressure monitoring device according to a modified British Hypertension Society protocol
Abstract
Background: The market for devices for the self-measurement of blood pressure is growing, and as accuracy is of prime importance, there is increasing pressure for manufacturers to provide evidence of independent testing. Recent reviews have shown that only five automated upper arm devices for self-measurement of blood pressure have been recommended for use. We tested the Microlife BP 3BTO-A, a lightweight, upper arm, automated oscillometric device, according to a modified version of the British Hypertension Society protocol and also analysed the computer-generated oscillograms for possible causes of inaccuracy.
Methods: One hundred and twenty-six subjects were recruited from general medical and specialist clinics and from amongst the staff at Guy's and St Thomas' Hospital, London, UK. Only 85 of these were included in the final analysis. Nine sequential readings were taken by two trained observers alternating between the mercury sphygmomanometer and the device. The last seven readings were analysed according to the British Hypertension Society protocol. Modifications to the protocol were: (1) the exclusion of patients whose blood pressure varied by more than 15 mmHg between sequential observer readings and (2) limited testing in the low systolic pressure range.
Results: The Microlife achieved a grade A for both systolic and diastolic pressure according to the British Hypertension Society protocol. The mean differences (standard deviation) between the observers and the device were -1.6 (7.7) mmHg and -2.1 (6.3) mmHg for systolic and diastolic blood pressure, respectively, therefore also fulfilling the criteria set by the Association for the Advancement of Medical Instrumentation. Sub-analysis for different pressure ranges showed that the device was less accurate in the high-pressure range (>160/100 mmHg).
Conclusion: The Microlife can be recommended for clinical use in an adult population.
Copyright 2002 Lippincott Williams & Wilkins
Similar articles
-
Validation of the Omron 705IT (HEM-759-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol.Blood Press Monit. 2006 Feb;11(1):27-32. doi: 10.1097/01.mbp.0000189788.05736.5f. Blood Press Monit. 2006. PMID: 16410738
-
Validation of the Omron M7 (HEM-780-E) oscillometric blood pressure monitoring device according to the British Hypertension Society protocol.Blood Press Monit. 2008 Feb;13(1):49-54. doi: 10.1097/MBP.0b013e3282cb57b6. Blood Press Monit. 2008. PMID: 18199924
-
An accurate automated blood pressure device for use in pregnancy and pre-eclampsia: the Microlife 3BTO-A.BJOG. 2005 Jul;112(7):915-20. doi: 10.1111/j.1471-0528.2005.00617.x. BJOG. 2005. PMID: 15957992
-
Oscillometric blood pressure measurement: progress and problems.Blood Press Monit. 2001 Dec;6(6):287-90. doi: 10.1097/00126097-200112000-00004. Blood Press Monit. 2001. PMID: 12055403 Review.
-
Can an automatic oscillometric device replace a mercury sphygmomanometer on blood pressure measurement? a systematic review and meta-analysis.Blood Press Monit. 2019 Dec;24(6):265-276. doi: 10.1097/MBP.0000000000000412. Blood Press Monit. 2019. PMID: 31658107
Cited by
-
Sources of inaccuracy in the measurement of adult patients' resting blood pressure in clinical settings: a systematic review.J Hypertens. 2017 Mar;35(3):421-441. doi: 10.1097/HJH.0000000000001197. J Hypertens. 2017. PMID: 27977471 Free PMC article.
-
Metabolic characteristics of ischaemic preconditioning induced performance improvement in Taekwondo athletes using LC‒MS/MS-based plasma metabolomics.Sci Rep. 2024 Oct 19;14(1):24609. doi: 10.1038/s41598-024-76045-1. Sci Rep. 2024. PMID: 39427043 Free PMC article.
-
Non-periodized and Daily Undulating Periodized Resistance Training on Blood Pressure of Older Women.Front Physiol. 2018 Nov 27;9:1525. doi: 10.3389/fphys.2018.01525. eCollection 2018. Front Physiol. 2018. PMID: 30542289 Free PMC article.
-
Sustained effect of resistance training on blood pressure and hand grip strength following a detraining period in elderly hypertensive women: a pilot study.Clin Interv Aging. 2014 Jan 20;9:219-25. doi: 10.2147/CIA.S56058. eCollection 2014. Clin Interv Aging. 2014. PMID: 24477221 Free PMC article. Clinical Trial.
-
Sarcopenia Is Associated with High Pulse Pressure in Older Women.J Aging Res. 2015;2015:109824. doi: 10.1155/2015/109824. Epub 2015 Aug 5. J Aging Res. 2015. PMID: 26346157 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources