Comparison of forced-air warming systems with lower body blankets using a copper manikin of the human body
- PMID: 12492798
- DOI: 10.1034/j.1399-6576.2003.470110.x
Comparison of forced-air warming systems with lower body blankets using a copper manikin of the human body
Abstract
Background: Forced-air warming has gained high acceptance as a measure for the prevention of intraoperative hypothermia. However, data on heat transfer with lower body blankets are not yet available. This study was conducted to determine the heat transfer efficacy of six complete lower body warming systems.
Methods: Heat transfer of forced-air warmers can be described as follows:[1]Qdot;=h.DeltaT.A where Qdot; = heat transfer [W], h = heat exchange coefficient [W m-2 degrees C-1], DeltaT = temperature gradient between blanket and surface [ degrees C], A = covered area [m2]. We tested the following forced-air warmers in a previously validated copper manikin of the human body: (1) Bair Hugger and lower body blanket (Augustine Medical Inc., Eden Prairie, MN); (2) Thermacare and lower body blanket (Gaymar Industries, Orchard Park, NY); (3) WarmAir and lower body blanket (Cincinnati Sub-Zero Products, Cincinnati, OH); (4) Warm-Gard(R) and lower body blanket (Luis Gibeck AB, Upplands Väsby, Sweden); (5) Warm-Gard and reusable lower body blanket (Luis Gibeck AB); and (6) WarmTouch and lower body blanket (Mallinckrodt Medical Inc., St. Luis, MO). Heat flux and surface temperature were measured with 16 calibrated heat flux transducers. Blanket temperature was measured using 16 thermocouples. DeltaT was varied between -10 and +10 degrees C and h was determined by a linear regression analysis as the slope of DeltaT vs. heat flux. Mean DeltaT was determined for surface temperatures between 36 and 38 degrees C, because similar mean skin temperatures have been found in volunteers. The area covered by the blankets was estimated to be 0.54 m2.
Results: Heat transfer from the blanket to the manikin was different for surface temperatures between 36 degrees C and 38 degrees C. At a surface temperature of 36 degrees C the heat transfer was higher (between 13.4 W to 18.3 W) than at surface temperatures of 38 degrees C (8-11.5 W). The highest heat transfer was delivered by the Thermacare system (8.3-18.3 W), the lowest heat transfer was delivered by the Warm-Gard system with the single use blanket (8-13.4 W). The heat exchange coefficient varied between 12.5 W m-2 degrees C-1 and 30.8 W m-2 degrees C-1, mean DeltaT varied between 1.04 degrees C and 2.48 degrees C for surface temperatures of 36 degrees C and between 0.50 degrees C and 1.63 degrees C for surface temperatures of 38 degrees C.
Conclusion: No relevant differences in heat transfer of lower body blankets were found between the different forced-air warming systems tested. Heat transfer was lower than heat transfer by upper body blankets tested in a previous study. However, forced-air warming systems with lower body blankets are still more effective than forced-air warming systems with upper body blankets in the prevention of perioperative hypothermia, because they cover a larger area of the body surface.
Similar articles
-
Comparison of forced-air warming systems with upper body blankets using a copper manikin of the human body.Acta Anaesthesiol Scand. 2002 Sep;46(8):965-72. doi: 10.1034/j.1399-6576.2002.460807.x. Acta Anaesthesiol Scand. 2002. PMID: 12190797
-
Efficacy of forced-air warming systems with full body blankets.Can J Anaesth. 2007 Jan;54(1):34-41. doi: 10.1007/BF03021897. Can J Anaesth. 2007. PMID: 17197466
-
Differences among forced-air warming systems with upper body blankets are small. A randomized trial for heat transfer in volunteers.Acta Anaesthesiol Scand. 2003 Oct;47(9):1159-64. doi: 10.1034/j.1399-6576.2003.00222.x. Acta Anaesthesiol Scand. 2003. PMID: 12969112 Clinical Trial.
-
Forced-air warming: technology, physical background and practical aspects.Curr Opin Anaesthesiol. 2009 Dec;22(6):769-74. doi: 10.1097/ACO.0b013e328331d134. Curr Opin Anaesthesiol. 2009. PMID: 19734783 Review.
-
Prevention of perioperative hypothermia with forced-air warming systems and upper-body blankets.Surg Technol Int. 2006;15:19-22. Surg Technol Int. 2006. PMID: 17029156 Review.
Cited by
-
Effect of forced-air warming by an underbody blanket on end-of-surgery hypothermia: a propensity score-matched analysis of 5063 patients.BMC Anesthesiol. 2019 Apr 9;19(1):50. doi: 10.1186/s12871-019-0724-8. BMC Anesthesiol. 2019. PMID: 30967133 Free PMC article.
-
Comparison of upper and lower body forced air blanket to prevent perioperative hypothermia in patients who underwent spinal surgery in prone position: a randomized controlled trial.Korean J Anesthesiol. 2022 Feb;75(1):37-46. doi: 10.4097/kja.21087. Epub 2021 May 13. Korean J Anesthesiol. 2022. PMID: 33984220 Free PMC article. Clinical Trial.
-
Heated humidified high-flow nasal oxygen prevents intraoperative body temperature decrease in non-intubated thoracoscopy.J Anesth. 2018 Dec;32(6):872-879. doi: 10.1007/s00540-018-2567-8. Epub 2018 Oct 15. J Anesth. 2018. PMID: 30324221
-
The Effect of Intraoperative Hypothermia on Anastomotic Leakage After Esophagectomy.Cancers (Basel). 2025 Mar 30;17(7):1166. doi: 10.3390/cancers17071166. Cancers (Basel). 2025. PMID: 40227724 Free PMC article.
-
What is the predictor of the intraoperative body temperature in abdominal surgery?J Anesth. 2019 Feb;33(1):67-73. doi: 10.1007/s00540-018-2585-6. Epub 2018 Nov 29. J Anesth. 2019. PMID: 30498930
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources