. 2003 Jan;34(1):77-83.
doi: 10.1161/01.str.0000046455.51363.e2.

Contrast burst depletion imaging (CODIM): a new imaging procedure and analysis method for semiquantitative ultrasonic perfusion imaging

Affiliations

Contrast burst depletion imaging (CODIM): a new imaging procedure and analysis method for semiquantitative ultrasonic perfusion imaging

Jens Eyding et al. Stroke. 2003 Jan.

Abstract

Background and purpose: Established methods of ultrasonic perfusion imaging using a bolus application of echo contrast agent provide only qualitative data because of various physical phenomena. This study was intended to investigate whether a new ultrasound perfusion imaging method termed contrast burst depletion imaging (CODIM) may provide semiquantitative measures of parenchymal perfusion independent of examination depth and acoustic energy distribution.

Methods: In a system with a constant concentration of contrast agent, analyzing the decrease in image intensity that occurs with microbubble-destructive imaging modes yields parameters that are considered to correlate with tissue perfusion. This method was first evaluated with a perfusion model that showed that the main resulting parameter "perfusion coefficient" (PC) is a monotonic nonlinear function of flow velocity. Seventeen human volunteers were then scanned according to this method with the use of 2 different contrast agents. Results were correlated with those from perfusion-weighted MRI examinations.

Results: The PC did not show significant differences in gray matter areas (ranging from 1.466x10(-2) x s(-1) to 1.641x10(-2) x s(-1)) of the brain despite different insonation depths (eg, ipsilateral and contralateral thalamus). In contrast, white matter exhibited significantly lower perfusion values in both imaging modes (PC: 0.604x10(-2) x s(-1) to 0.745x10(-2) x s(-1); P<0.05).

Conclusions: CODIM is a promising new tool of imaging parenchymal (brain) perfusion in healthy persons. The method provides semiquantitative and depth-independent perfusion parameters and in this way overcomes the limitations of the perfusion methods using a bolus kinetic. Further investigations must be done to evaluate the potential of the method in patients with perfusion deficits.

PubMed Disclaimer