. 2002 Dec;29(12):2767-81.
doi: 10.1118/1.1523932.

A CdZnTe slot-scanned detector for digital mammography

Affiliations

A CdZnTe slot-scanned detector for digital mammography

James G Mainprize et al. Med Phys. 2002 Dec.

Abstract

A new high-resolution detector has been developed for use in a slot-scanned digital mammography system. The detector is a hybrid device that consists of a CCD operating in time-delay integration mode that is bonded to a 150-microm-thick CdZnTe photoconductor array. The CCD was designed with a detector element pitch of 50 microm. Two devices were evaluated with differing crystalline quality. Incomplete charge collection was a source of reduction in DQE. This occurs in both devices due to characteristically low mobility-lifetime products for CdZnTe, with the greatest losses demonstrated by the multicrystalline sample. The mobility-lifetime products for the multicrystalline device were found to be 2.4 x 10(-4) and 4.0 x 10(-7) cm2/V for electrons and holes, respectively. The device constructed with higher quality single crystal CdZnTe demonstrated mobility-lifetime products of 1.0 x 10(-4) and 4.4 x 10(-6) cm2/V for electrons and holes. The MTF and DQE for the device were measured at several exposures and results were compared to predictions from a linear systems model of signal and noise propagation. The MTF at a spatial frequency of 10 mm(-1) exceeded 0.18 and 0.56 along the scan and slot directions, respectively. Scanning motion and CCD design limited the resolution along the scan direction. For an x-ray beam from a tungsten target tube with 40 microm molybdenum filtration operated at 26 kV, the single crystal device demonstrated a DQE(0) of 0.70 +/- 0.02 at 7.1 x 10(-6) C/kg (27 mR) exposure to the detector, despite its relatively poor charge collection efficiency.

PubMed Disclaimer

LinkOut - more resources