A comparison of ex vivo and in vitro Sutter metacarpophalangeal prostheses
- PMID: 12531676
- DOI: 10.1054/jhsb.2002.0862
A comparison of ex vivo and in vitro Sutter metacarpophalangeal prostheses
Abstract
Forty-one Sutter metacarpophalangeal prostheses were implanted into 11 hands of nine patients. Twelve of these prostheses were revised from three patients after a mean period of 42 months. Of the 12 prostheses, 11 showed fracture at the junction of the distal stem and the central hinge region. Two Sutter metacarpophalangeal prostheses were tested on a single station finger simulator and both failed due to fracture at the junction of the hinge and the distal stem.
Comment in
-
A comparison of ex vivo and in vitro Sutter metacarpophalangeal prostheses, 28B: 86-91.J Hand Surg Br. 2003 Oct;28(5):505; author reply 505-6. doi: 10.1016/s0266-7681(03)00166-9. J Hand Surg Br. 2003. PMID: 14562795 No abstract available.