8.0-Tesla human MR system: temperature changes associated with radiofrequency-induced heating of a head phantom
- PMID: 12541230
- DOI: 10.1002/jmri.10236
8.0-Tesla human MR system: temperature changes associated with radiofrequency-induced heating of a head phantom
Abstract
Purpose: To investigate if the heat induced in biological tissues by typical radio frequency (RF) energy associated with an 8.0-Tesla magnetic resonance (MR) system causes excessive temperature changes.
Materials and methods: Fluoroptic thermometry was used to measure temperatures in multiple positions in a head phantom made of ground turkey breast. A series of experiments were conducted with measurements obtained at RF power levels ranging from a specific absorption rate (SAR) of up to 4.0 W/kg for 10 minutes.
Results: The highest temperature increases were up to 0.7 degrees C. An inhomogeneous heating pattern was observed. In general, the deep regions within the phantom registered higher temperature increases compared to the peripheral sites.
Conclusion: The expectation of an inhomogeneous RF distribution in ultra high field systems (> 4 T) was confirmed. At a frequency of 340 MHz and in-tissue RF wave length of about 10 cm, the RF inhomogeneity was measured to create higher temperatures in deeper regions of a human head phantom compared to peripheral tissues. Our results agree with the computational electromagnetic calculations for such frequencies. Importantly, these experiments indicated that there were no regions of heating that exceeded the current FDA guidelines.
Copyright 2003 Wiley-Liss, Inc.
Similar articles
-
Effects of coil dimensions and field polarization on RF heating inside a head phantom.Magn Reson Imaging. 2005 Jan;23(1):53-60. doi: 10.1016/j.mri.2004.11.007. Magn Reson Imaging. 2005. PMID: 15733788
-
Variability in RF-induced heating of a deep brain stimulation implant across MR systems.J Magn Reson Imaging. 2006 Dec;24(6):1236-42. doi: 10.1002/jmri.20769. J Magn Reson Imaging. 2006. PMID: 17078088
-
Evaluation of specific absorption rate as a dosimeter of MRI-related implant heating.J Magn Reson Imaging. 2004 Aug;20(2):315-20. doi: 10.1002/jmri.20103. J Magn Reson Imaging. 2004. PMID: 15269959
-
Ocular effects of radiofrequency energy.Bioelectromagnetics. 2003;Suppl 6:S148-61. doi: 10.1002/bem.10117. Bioelectromagnetics. 2003. PMID: 14628311 Review.
-
MRI-induced heating of selected thin wire metallic implants-- laboratory and computational studies-- findings and new questions raised.Minim Invasive Ther Allied Technol. 2006;15(2):76-84. doi: 10.1080/13645700600640931. Minim Invasive Ther Allied Technol. 2006. PMID: 16754190 Review.
Cited by
-
Magnetic resonance safety.Abdom Radiol (NY). 2016 Mar;41(3):444-51. doi: 10.1007/s00261-016-0680-4. Abdom Radiol (NY). 2016. PMID: 26940331 Free PMC article. Review.
-
Radiofrequency heating at 9.4T: in vivo temperature measurement results in swine.Magn Reson Med. 2008 Jan;59(1):73-8. doi: 10.1002/mrm.21425. Magn Reson Med. 2008. PMID: 17969077 Free PMC article.
-
Whole-body MRI at high field: technical limits and clinical potential.Eur Radiol. 2005 May;15(5):946-59. doi: 10.1007/s00330-005-2678-0. Epub 2005 Jan 27. Eur Radiol. 2005. PMID: 15856252 Review.
-
Biological effects of exposure to magnetic resonance imaging: an overview.Biomed Eng Online. 2004 Apr 22;3:11. doi: 10.1186/1475-925X-3-11. Biomed Eng Online. 2004. PMID: 15104797 Free PMC article. Review.
-
Radiofrequency heating in porcine models with a "large" 32 cm internal diameter, 7 T (296 MHz) head coil.Magn Reson Med. 2011 Jul;66(1):255-63. doi: 10.1002/mrm.22790. Epub 2011 Feb 17. Magn Reson Med. 2011. PMID: 21337423 Free PMC article.