Fiberglass needle electrodes for transmural cardiac mapping
- PMID: 12549747
- DOI: 10.1109/TBME.2002.805483
Fiberglass needle electrodes for transmural cardiac mapping
Abstract
We developed a new method for fabricating plunge needle electrodes for use in cardiac mapping. The needles have 12 electrodes with 1-mm spacing, are 0.5 mm in diameter, and are fabricated from fiberglass reinforced epoxy. They are stiff enough to be easily inserted into beating hearts and durable enough to be reused many times. We found that these new needles elicit smaller, more quickly resolving injury potentials, and when inserted in a row with 2-mm spacing, disrupt ventricular fibrillation activation patterns less than traditional steel needles.
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources