The evaluation of tonometry and self-tonometry with Ocuton tonometers
- PMID: 12552251
The evaluation of tonometry and self-tonometry with Ocuton tonometers
Abstract
Background: The purpose of the present study was to evaluate the clinical applicability of the Ocuton-A and Ocuton-S applanation tonometers, and to compare their use and accuracy parameters to those of Goldmann applanation tonometry (GAT).
Material/methods: In the first study, intraocular pressure (IOP) was measured with an Ocuton-A tonometer, followed by another measurement using GAT in 15 subjects (30 eyes). Ocuton-tonometry was performed on three occasions separated by three-minute intervals. In Study 2 the impact of increased patient familiarity on measurement accuracy using the self-tonometer was investigated in 5 subjects (10 eyes) by comparing two consecutive series of five Ocuton-S measurements. In Study 3 nine trained volunteers measured their own IOP with an Ocuton-S self-tonometer after GAT and Ocuton-A measurements by an experienced investigator.
Results: Study 1 showed that the measured IOP value was significantly higher using the Ocuton-A tonometer, compared to values obtained using GAT. In Study 2 we found that increased practice in self-tonometry did not alter measurement accuracy. In Study 3 the difference between IOP readings obtained with the Ocuton-S and Ocuton-A instruments was not significant.
Conclusions: The portable Ocuton-A tonometer may become a useful instrument for IOP measurement by professional workers outside the clinic. Qualified patients are able to use the Ocuton-S self-tonometer reliably even after limited training, and its measurement accuracy is acceptable for IOP monitoring. Both the Ocuton A and S devices consistently overestimate the IOP by approximately 2 mm Hg compared to the corresponding Goldmann readings.
Similar articles
-
Intraocular pressure measurements with the newly reconfigured Ocuton S*TT-MV self-tonometer in comparison to Goldmann applanation tonometry in glaucoma patients.Med Sci Monit. 2009 Nov;15(11):CR556-62. Med Sci Monit. 2009. PMID: 19865054
-
Self-tonometry with the Ocuton S: evaluation of accuracy in glaucoma patients.Acta Ophthalmol Scand. 2004 Aug;82(4):405-9. doi: 10.1111/j.1600-0420.2004.00279.x. Acta Ophthalmol Scand. 2004. PMID: 15291933
-
[Intraocular pressure measurements with the Proview self-tonometer in comparison of Goldmann applanation tonometry in healthy and glaucomatous eyes].Klin Monbl Augenheilkd. 2006 Nov;223(11):899-903. doi: 10.1055/s-2006-927094. Klin Monbl Augenheilkd. 2006. PMID: 17131250 Clinical Trial. German.
-
[What to do if the intraocular pressure measurement does not appear reliable].J Fr Ophtalmol. 2010 Apr;33(4):279-84. doi: 10.1016/j.jfo.2010.02.007. Epub 2010 Mar 29. J Fr Ophtalmol. 2010. PMID: 20347507 Review. French.
-
Systematic review of the agreement of tonometers with Goldmann applanation tonometry.Ophthalmology. 2012 Aug;119(8):1552-7. doi: 10.1016/j.ophtha.2012.02.030. Epub 2012 May 10. Ophthalmology. 2012. PMID: 22578443
Cited by
-
Relationship of systemic blood pressure with ocular perfusion pressure and intraocular pressure of glaucoma patients in telemedical home monitoring.Med Sci Monit. 2012 Nov;18(11):MT85-9. doi: 10.12659/msm.883529. Med Sci Monit. 2012. PMID: 23111749 Free PMC article. Clinical Trial.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources