Stemmed femoral knee prostheses: effects of prosthetic design and fixation on bone loss
- PMID: 12553509
- DOI: 10.1080/000164702321039589
Stemmed femoral knee prostheses: effects of prosthetic design and fixation on bone loss
Abstract
Although the revision rates for modern knee prostheses have decreased drastically, the total number of revisions a year is increasing because many more primary knee replacements are being done. At the time of revision, bone loss is common, which compromises prosthetic stability. To improve stability, intramedullary stems are often used. The aim of this study was to estimate the effects of a stem, its diameter and the interface bonding conditions on patterns of the bone remodeling in the distal femur. We created finite element models of the distal half of a femur in which 4 types of knee prostheses were placed. The bone remodeling process was simulated using a strain-adaptive bone remodeling theory. The amount of such remodeling was determined by calculating the changes in bone mineral density in 9 regions of interest from simulated DEXA scans. The computer simulation model showed that revision prostheses tend to cause more bone resorption than primary ones, especially in the most distal regions. Predicted long-term bone loss due to a revision prosthesis with a thin stem equalled that around a prosthesis with an intercondylar box. However, strong regional differences were found--the stemmed prostheses having more bone loss in the most distal areas and some bone gain in the more proximal ones. A prosthesis with a thick stem led to an increase in bone loss. When the prosthesis-cement interface was bonded, more bone loss was predicted than with an unbonded interface. These results suggest that a stem which increases stability initially may reduce stability in the long term. This is due to an increase in stress shielding and bone resorption.
Similar articles
-
Effects of material properties of femoral hip components on bone remodeling.J Orthop Res. 1992 Nov;10(6):845-53. doi: 10.1002/jor.1100100614. J Orthop Res. 1992. PMID: 1403299
-
Preclinical testing of total hip stems. The effects of coating placement.Clin Orthop Relat Res. 1995 Oct;(319):64-76. Clin Orthop Relat Res. 1995. PMID: 7554651
-
The relationship between stress shielding and bone resorption around total hip stems and the effects of flexible materials.Clin Orthop Relat Res. 1992 Jan;(274):124-34. Clin Orthop Relat Res. 1992. PMID: 1728998
-
Determinants of stress shielding: design versus materials versus interface.Clin Orthop Relat Res. 1992 Jan;(274):202-12. Clin Orthop Relat Res. 1992. PMID: 1729005 Review.
-
[The causes of failure for hip and knee arthroplasties].Ned Tijdschr Geneeskd. 1998 Sep 12;142(37):2035-40. Ned Tijdschr Geneeskd. 1998. PMID: 9856208 Review. Dutch.
Cited by
-
A sliding stem in revision total knee arthroplasty provides stability and reduces stress shielding.Acta Orthop. 2010 Jun;81(3):337-43. doi: 10.3109/17453674.2010.483991. Acta Orthop. 2010. PMID: 20450422 Free PMC article.
-
The influence of body weight index on initial stability of uncemented femoral knee protheses: A finite element study.Heliyon. 2023 Feb 17;9(3):e13819. doi: 10.1016/j.heliyon.2023.e13819. eCollection 2023 Mar. Heliyon. 2023. PMID: 36895366 Free PMC article.
-
Densitometric evaluation of bone-prosthetic counterface in hip and knee arthroplasty with modern implants.Clin Cases Miner Bone Metab. 2016 May-Aug;13(2):144-150. doi: 10.11138/ccmbm/2016.13.2.144. Epub 2016 Oct 5. Clin Cases Miner Bone Metab. 2016. PMID: 27920813 Free PMC article.
-
Load-induced deformation of the tibia and its effect on implant loosening detection.Sci Rep. 2023 Dec 8;13(1):21769. doi: 10.1038/s41598-023-49177-z. Sci Rep. 2023. PMID: 38066256 Free PMC article.
-
Influence of a metaphyseal sleeve on the stress-strain state of a bone-tumor implant system in the distal femur: an experimental and finite element analysis.J Orthop Surg Res. 2020 Dec 9;15(1):589. doi: 10.1186/s13018-020-02025-6. J Orthop Surg Res. 2020. PMID: 33298115 Free PMC article.