Artifacts at PET and PET/CT caused by metallic hip prosthetic material
- PMID: 12563158
- DOI: 10.1148/radiol.2262012141
Artifacts at PET and PET/CT caused by metallic hip prosthetic material
Abstract
Hip prosthetic material and a steel rod were scanned in a water bath of fluorine 18 fluorodeoxyglucose (FDG) with positron emission tomographic (PET) and PET/computed tomographic (CT) scanners to evaluate the generation of artifacts adjacent to the metal. The influences of attenuation correction (AC), positioning of the object, and image reconstruction were examined. Use of CT- and germanium 68-based AC resulted in generation of artifacts that mimicked increased FDG uptake. These artifacts were more evident when the object was moved between the emission and transmission scans. When attenuation-weighted iterative reconstruction was used, these artifacts were less evident.
Similar articles
-
Value of iterative reconstruction, attenuation correction, and image fusion in the interpretation of FDG PET images with an integrated dual-head coincidence camera and X-ray-based attenuation maps.Radiology. 2001 Jan;218(1):163-71. doi: 10.1148/radiology.218.1.r01ja59163. Radiology. 2001. PMID: 11152796 Clinical Trial.
-
A virtual sinogram method to reduce dental metallic implant artefacts in computed tomography-based attenuation correction for PET.Nucl Med Commun. 2010 Jan;31(1):22-31. doi: 10.1097/MNM.0b013e32832fa241. Nucl Med Commun. 2010. PMID: 19829166
-
X-ray-based attenuation correction for positron emission tomography/computed tomography scanners.Semin Nucl Med. 2003 Jul;33(3):166-79. doi: 10.1053/snuc.2003.127307. Semin Nucl Med. 2003. PMID: 12931319 Review.
-
Do hardware artefacts influence the performance of head and neck PET scans in patients with oral cavity squamous cell cancer?Dentomaxillofac Radiol. 2003 Nov;32(6):365-71. doi: 10.1259/dmfr/77741718. Dentomaxillofac Radiol. 2003. PMID: 15070838
-
Positron emission tomography/computed tomography potential pitfalls and artifacts.Curr Probl Diagn Radiol. 2009 Jul-Aug;38(4):156-69. doi: 10.1067/j.cpradiol.2008.01.001. Curr Probl Diagn Radiol. 2009. PMID: 19464586 Review.
Cited by
-
Assessment of the diagnostic accuracy of 18F-FDG PET/CT in prosthetic infective endocarditis and cardiac implantable electronic device infection: comparison of different interpretation criteria.Eur J Nucl Med Mol Imaging. 2016 Dec;43(13):2401-2412. doi: 10.1007/s00259-016-3463-9. Epub 2016 Sep 5. Eur J Nucl Med Mol Imaging. 2016. PMID: 27596984 Clinical Trial.
-
Diagnostic value of 18F-FDG PET/CT in trauma patients with suspected chronic osteomyelitis.Eur J Nucl Med Mol Imaging. 2007 May;34(5):704-714. doi: 10.1007/s00259-006-0290-4. Epub 2006 Nov 29. Eur J Nucl Med Mol Imaging. 2007. PMID: 17136411
-
Application of 68Ga-citrate PET/CT for differentiating periprosthetic joint infection from aseptic loosening after joint replacement surgery.Bone Joint Res. 2022 Jun;11(6):398-408. doi: 10.1302/2046-3758.116.BJR-2021-0464.R1. Bone Joint Res. 2022. PMID: 35731211 Free PMC article.
-
Accuracy of digital model generated from CT data with metal artifact reduction algorithm.Sci Rep. 2021 May 14;11(1):10332. doi: 10.1038/s41598-021-89298-x. Sci Rep. 2021. PMID: 33990637 Free PMC article.
-
Assessment of errors caused by X-ray scatter and use of contrast medium when using CT-based attenuation correction in PET.Eur J Nucl Med Mol Imaging. 2006 Nov;33(11):1301-13. doi: 10.1007/s00259-006-0086-6. Epub 2006 Apr 19. Eur J Nucl Med Mol Imaging. 2006. PMID: 16622710
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical