Respiratory gating for liver tumors: use in dose escalation
- PMID: 12573753
- DOI: 10.1016/s0360-3016(02)03941-x
Respiratory gating for liver tumors: use in dose escalation
Abstract
Purpose: To determine the clinical impact of the Varian Real-Time Position Monitor (RPM) respiratory gating system for treatment of liver tumors.
Methods and materials: Ten patients with liver tumors were selected for evaluation of this passive system, which tracks motion of reflective markers mounted on the abdomen with an infrared-sensitive camera. At simulation, a fluoroscopic movie, breathing trace, and CT scans synchronized at end-expiration (E-E) and end-inspiration were acquired in treatment position using the RPM system. Organs and gross tumor volume were contoured on each CT. Each organ's positional change between two scan sets was quantified by calculation of the center of volume shift and an "index coefficient," defined as the volume common to the two versions of the organ to the volume included in at least one (intersection/union). Treatment dose was determined by use of normal tissue complication probability calculations and dose-volume histograms. Gated portal images were obtained to monitor gating reproducibility with treatment.
Results: Eight patients received 177 treatments with RPM gating. Average superior-to-inferior (SI) diaphragm motion on initial fluoroscopy was reduced from 22.7 mm without gating to 5.1 mm with gating. Comparing end-inspiration to E-E CT scans, average SI movement of the right diaphragm was 11.5 mm vs. 2.2 mm for two E-E CT scans. For all organs, average E-I SI organ motion was 12.8 mm vs. 2.0 mm for E-E studies. Index coefficients were closer to 1.0 for E-E than end-inspiration scans, indicating gating reproducibility. The average SI displacement of diaphragm apex on gated portal images compared with DRR was 2.3 mm. Treatment was prolonged less than 10 minutes with gating. The reproducible decrease in organ motion with gating enabled reduction in gross tumor volume-to-planning target volume margin from 2 to 1 cm. This allowed for calculated dose increases of 7%-27% (median: 21.3%) in 6 patients and enabled treatment in 2.
Conclusion: Gating of radiotherapy for liver tumors enables safe margin reduction on tumor volume, which, in turn, may allow for dose escalation.
Similar articles
-
Reduction of organ motion in lung tumors with respiratory gating.Lung Cancer. 2006 Jan;51(1):41-51. doi: 10.1016/j.lungcan.2005.08.008. Epub 2005 Sep 28. Lung Cancer. 2006. PMID: 16198022
-
Deep inspiration breath-hold technique for lung tumors: the potential value of target immobilization and reduced lung density in dose escalation.Int J Radiat Oncol Biol Phys. 1999 Oct 1;45(3):603-11. doi: 10.1016/s0360-3016(99)00154-6. Int J Radiat Oncol Biol Phys. 1999. PMID: 10524412
-
Dosimetric analysis of respiratory-gated radiotherapy for hepatocellular carcinoma.Med Dosim. 2011 Summer;36(2):213-8. doi: 10.1016/j.meddos.2010.03.006. Epub 2010 Jun 1. Med Dosim. 2011. PMID: 20570128
-
Evaluation of respiratory movement during gated radiotherapy using film and electronic portal imaging.Int J Radiat Oncol Biol Phys. 2002 Feb 1;52(2):522-31. doi: 10.1016/s0360-3016(01)02681-5. Int J Radiat Oncol Biol Phys. 2002. PMID: 11872300
-
[Internal target volume definition using four-dimensional CT and dosimetric evaluation for hepatocellular carcinoma].Ai Zheng. 2007 Jan;26(1):1-8. Ai Zheng. 2007. PMID: 17222359 Chinese.
Cited by
-
Increased Risk of Diabetes after Definitive Radiotherapy in Patients with Indolent Gastroduodenal Lymphoma.Cancer Res Treat. 2022 Jan;54(1):294-300. doi: 10.4143/crt.2021.073. Epub 2021 Apr 1. Cancer Res Treat. 2022. PMID: 33794083 Free PMC article.
-
Are fiducial markers useful surrogates when using respiratory gating to reduce motion of gastroesophageal junction tumors?Acta Oncol. 2016 Aug;55(8):1040-6. doi: 10.3109/0284186X.2016.1167953. Epub 2016 May 6. Acta Oncol. 2016. PMID: 27152887 Free PMC article. Clinical Trial.
-
Gated Volumetric-Modulated Arc Therapy vs. Tumor-Tracking CyberKnife Radiotherapy as Stereotactic Body Radiotherapy for Hepatocellular Carcinoma: A Dosimetric Comparison Study Focused on the Impact of Respiratory Motion Managements.PLoS One. 2016 Nov 22;11(11):e0166927. doi: 10.1371/journal.pone.0166927. eCollection 2016. PLoS One. 2016. PMID: 27875568 Free PMC article.
-
Management of three-dimensional intrafraction motion through real-time DMLC tracking.Med Phys. 2008 May;35(5):2050-61. doi: 10.1118/1.2905355. Med Phys. 2008. PMID: 18561681 Free PMC article.
-
Helical tomotherapy for single and multiple liver tumours.Radiat Oncol. 2010 Jun 24;5:58. doi: 10.1186/1748-717X-5-58. Radiat Oncol. 2010. PMID: 20576108 Free PMC article.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous