Discrepancies between catheter and Doppler estimates of valve effective orifice area can be predicted from the pressure recovery phenomenon: practical implications with regard to quantification of aortic stenosis severity
- PMID: 12575972
- DOI: 10.1016/s0735-1097(02)02764-x
Discrepancies between catheter and Doppler estimates of valve effective orifice area can be predicted from the pressure recovery phenomenon: practical implications with regard to quantification of aortic stenosis severity
Abstract
Objectives: We sought to obtain more coherent evaluations of aortic stenosis severity.
Background: The valve effective orifice area (EOA) is routinely used to assess aortic stenosis severity. However, there are often discrepancies between measurements of EOA by Doppler echocardiography (EOA(Dop)) and those by a catheter (EOA(cath)). We hypothesized that these discrepancies might be due to the influence of pressure recovery.
Methods: The relationship between EOA(cath) and EOA(Dop) was studied as follows: 1) in an in vitro model measuring the effects of different flow rates and aortic diameters on two fixed stenoses and seven bioprostheses; 2) in an animal model of supravalvular aortic stenosis (14 pigs); and 3) based on catheterization data from 37 patients studied by Schöbel et al.
Results: Pooling of in vitro, animal, and patient data showed a good correlation (r = 0.97) between EOA(cath) (range 0.3 to 2.3 cm(2)) and EOA(Dop) (range 0.2 to 1.7 cm(2)), but EOA(cath) systematically overestimated EOA(Dop) (24 +/- 17% [mean +/- SD]). However, when the energy loss coefficient (ELCo) was calculated from EOA(Dop) and aortic cross-sectional area (A(A)) to account for pressure recovery, a similar correlation (r = 0.97) with EOA(cath) was observed, but the previously noted overestimation was no longer present.
Conclusions: Discrepancies between EOA(cath) and EOA(Dop) are largely due to the pressure recovery phenomenon and can be reconciled by calculating ELCo from the echocardiogram. Thus, ELCo and EOA(cath) are equivalent indexes representing the net energy loss due to stenosis and probably are the most appropriate for quantifying aortic stenosis severity.
Comment in
-
Stenosis is in the eye of the observer: impact of pressure recovery on assessing aortic valve area.J Am Coll Cardiol. 2003 Feb 5;41(3):443-5. doi: 10.1016/s0735-1097(02)02765-1. J Am Coll Cardiol. 2003. PMID: 12575973 No abstract available.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources