A comparison of pulse oximetry and near infrared spectroscopy (NIRS) in the detection of hypoxaemia occurring with pauses in nasal airflow in neonates

Affiliations
Clinical Trial

A comparison of pulse oximetry and near infrared spectroscopy (NIRS) in the detection of hypoxaemia occurring with pauses in nasal airflow in neonates

S L Watkin et al. J Clin Monit Comput. 1999 Dec.

Abstract

Objective: The aim of this study was to compare the ability of NIRS and pulse oximetry to detect changes in cerebral oxygenation occurring in response to a pause in nasal airflow (PNA).

Methods: Twenty-one recordings of cerebral oxygenation index by NIRS together with oxyhemoglobin saturation by pulse oximetry were measured on 17 preterm infants with a history of apnoea. Photoplethysmography was used to confirm the accuracy of the pulse oximetry data. PNA events were defined as pauses of greater than 4 seconds in a thermistor trace measuring nasal air flow.

Results: Baseline variability in oxygenation index (Hbdiff) was found to be from -0.12 to +0.13 micromol 100 g brain(-1). A fall in Hbdiff or SpO2 was defined as a decrease of greater magnitude than 2 standard deviations from the baseline, i.e., -0.12 micromol 100 g brain(-1) and 3% respectively. In 68% of 468 PNA events a fall in oxyhemoglobin saturation (SpO2) was detected and in 56% a fall in Hbdiff was detected. In 20% of events there was no fall in cerebral oxygenation despite a fall in SpO2. In 8% of PNA episodes we recorded a fall in cerebral oxygenation but no fall in SpO2. When a fall in cerebral oxygenation was recorded, the fall was greater when the event was also associated with a fall in SpO2 (median (interquartile range (IQR)) 0.32 (0.21-0.69) vs. 0.25 (0.16-0.43) micromol 100 g brain(-1), p < 0.05). When all the PNA episodes were reviewed no close correlation was shown between the magnitude of change in cerebral oxygenation and the change in SpO2 for small changes in both indices. However, large falls (>1.5 micromol 100 g brain(-1)) in cerebral oxygenation were closely associated with large changes in SpO2.

Conclusions: We conclude that both techniques are sensitive to changes in oxygenation during PNA. Small changes in cerebral Hbdiff and arterial SpO2 do not always correlate for physiological reasons. A change in Hbdiff of >0.3 micromol 100 g brain(-1) is likely to be physiologically significant and is associated with a change in SpO2 of 12%.