A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques
- PMID: 12578081
- DOI: 10.1023/a:1009982611386
A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques
Abstract
Introduction: Bias and precision statistics have succeeded regression analysis when measurement techniques are compared. However, when applied to cardiac output measurements, inconsistencies occur in reporting the results of this form of analysis.
Methods: A MEDLINE search was performed, dating from 1986. Studies comparing techniques of cardiac output measurement using bias and precision statistics were surveyed. An error-gram was constructed from the percentage errors in the test and reference methods and was used to determine acceptable limits of agreement between methods.
Results: Twenty-five articles were found. Presentation of statistical data varied greatly. Four different statistical parameters were used to describe the agreement between measurements. The overall limits of agreement in studies evaluating bioimpedance (n = 23) was +/-37% (15-82%) and in those evaluating Doppler ultrasound (n = 11) +/-65% (25-225%). Objective criteria used to assess outcome were given in only 44% of the articles. These were (i) limits of agreement approaching +/-15-20%, (ii) limits of agreement of less than 1 L/min, and (iii) more than 75% of bias measurements within +/-20% of the mean. Graphically, we showed that limits of agreement of up to +/-30% were acceptable.
Conclusions: When using bias and precision statistics, cardiac output, bias, limits of agreement, and percentage error should be presented. Using current reference methods, acceptance of a new technique should rely on limits of agreement of up to +/-30%.
Similar articles
-
Comparing Methods for Cardiac Output: Intraoperatively Doppler-Derived Cardiac Output Measured With 3-Dimensional Echocardiography Is Not Interchangeable With Cardiac Output by Pulmonary Catheter Thermodilution.Anesth Analg. 2018 Aug;127(2):399-407. doi: 10.1213/ANE.0000000000002800. Anesth Analg. 2018. PMID: 29324489
-
Minimally invasive measurement of cardiac output during surgery and critical care: a meta-analysis of accuracy and precision.Anesthesiology. 2010 Nov;113(5):1220-35. doi: 10.1097/ALN.0b013e3181ee3130. Anesthesiology. 2010. PMID: 20881596 Review.
-
Evaluation of two methods for continuous cardiac output assessment during exercise in chronic heart failure patients.J Appl Physiol (1985). 2008 Dec;105(6):1822-9. doi: 10.1152/japplphysiol.90430.2008. Epub 2008 Oct 23. J Appl Physiol (1985). 2008. PMID: 18948448
-
Cardiac output determination from endotracheally measured impedance cardiography: clinical evaluation of endotracheal cardiac output monitor.J Cardiothorac Vasc Anesth. 2011 Oct;25(5):770-5. doi: 10.1053/j.jvca.2010.12.017. Epub 2011 Feb 26. J Cardiothorac Vasc Anesth. 2011. PMID: 21354828
-
Cardiac output measurements. A review of current techniques and research.Ann Biomed Eng. 1986;14(3):219-39. doi: 10.1007/BF02584272. Ann Biomed Eng. 1986. PMID: 3532871 Review.
Cited by
-
Agreement between continuous and intermittent pulmonary artery thermodilution for cardiac output measurement in perioperative and intensive care medicine: a systematic review and meta-analysis.Crit Care. 2021 Mar 29;25(1):125. doi: 10.1186/s13054-021-03523-7. Crit Care. 2021. PMID: 33781307 Free PMC article.
-
Cardiac output and cardiac index measured with cardiovascular magnetic resonance in healthy subjects, elite athletes and patients with congestive heart failure.J Cardiovasc Magn Reson. 2012 Jul 28;14(1):51. doi: 10.1186/1532-429X-14-51. J Cardiovasc Magn Reson. 2012. PMID: 22839436 Free PMC article.
-
Wearable Seismocardiography-Based Assessment of Stroke Volume in Congenital Heart Disease.J Am Heart Assoc. 2022 Sep 20;11(18):e026067. doi: 10.1161/JAHA.122.026067. Epub 2022 Sep 14. J Am Heart Assoc. 2022. PMID: 36102243 Free PMC article.
-
Pulse Wave Transit Time Measurements of Cardiac Output in Septic Shock Patients: A Comparison of the Estimated Continuous Cardiac Output System with Transthoracic Echocardiography.PLoS One. 2015 Jun 30;10(6):e0130489. doi: 10.1371/journal.pone.0130489. eCollection 2015. PLoS One. 2015. PMID: 26126112 Free PMC article.
-
Can arterial shape change with IABP play a role in comparing FlowTrac/EV 1000 vs. pulmonary cardiac output?Heart Vessels. 2024 Mar;39(3):273-274. doi: 10.1007/s00380-023-02346-6. Epub 2023 Dec 12. Heart Vessels. 2024. PMID: 38085331 No abstract available.