Autosynchronized systolic unloading during left ventricular assist with a centrifugal pump
- PMID: 12579105
- DOI: 10.1067/mtc.2003.100
Autosynchronized systolic unloading during left ventricular assist with a centrifugal pump
Abstract
Objectives: The purpose of this study was to investigate how the inflow cannulation site of the left ventricular assist system with a centrifugal pump would influence cardiac function on failing heart models.
Methods: In 10 sheep, a left ventricular assist system was instituted by an outflow cannula in the descending aorta, two inflow cannulas in the left atrium and the left ventricle, and connecting those cannulas to a magnetically suspended centrifugal pump. A conductance catheter and a tipped micromanometer for monitoring the pressure-volume loop were also inserted into the left ventricle. Myocardial oxygen consumption was directly measured. Heart failure was induced by injection of microspheres into the left main coronary artery. The assist rate was varied from 0% to 100% at each inflow cannulation site.
Results: The pump flow with left ventricular cannulation increased during the systolic phase and decreased during the diastolic phase, whereas it was constant with left atrial cannulation. Ejection fraction with left atrial cannulation decreased as the assist rate increased, whereas that with left ventricular cannulation was maintained up to 75% assist. The external work with left atrial cannulation decreased gradually as the assist rate increased, whereas the external work with left ventricular cannulation did not decrease until the assist rate reached 75%. The myocardial oxygen consumption in both cannulations decreased proportionally as the assist rate increased; they were significantly less with left ventricular cannulation at the 100% assist rate than with left atrial cannulation.
Conclusion: Left ventricular cannulation during left ventricular assistance maintains ejection fraction and effectively reduces oxygen consumption.
Similar articles
-
The impact of pump speed and inlet cannulation site on left ventricular unloading with a rotary blood pump.Artif Organs. 2004 Jul;28(7):660-7. doi: 10.1111/j.1525-1594.2004.07374.x. Artif Organs. 2004. PMID: 15209860
-
Effects of a left ventricular assist device with a centrifugal pump on left ventricular diastolic hemodynamics.Ann Thorac Cardiovasc Surg. 2002 Oct;8(5):275-80. Ann Thorac Cardiovasc Surg. 2002. PMID: 12472409
-
Ventricular actuation improves systolic and diastolic myocardial function in the small failing heart.Ann Thorac Surg. 2009 Dec;88(6):1982-8; discussion 1988. doi: 10.1016/j.athoracsur.2009.08.048. Ann Thorac Surg. 2009. PMID: 19932272
-
Augmentation of dilated failing left ventricular stroke work by a physiological cardiac assist device.Ann N Y Acad Sci. 2004 May;1015:379-90. doi: 10.1196/annals.1302.033. Ann N Y Acad Sci. 2004. PMID: 15201176 Review.
-
Regulation of coronary blood flow during exercise.Physiol Rev. 2008 Jul;88(3):1009-86. doi: 10.1152/physrev.00045.2006. Physiol Rev. 2008. PMID: 18626066 Review.
Cited by
-
The role of acute circulatory support in ST-segment elevation myocardial infarction complicated by cardiogenic shock.Indian Heart J. 2017 Sep-Oct;69(5):668-674. doi: 10.1016/j.ihj.2017.05.011. Epub 2017 May 22. Indian Heart J. 2017. PMID: 29054200 Free PMC article. Review.
-
Percutaneous Mechanical Circulatory Support for Cardiogenic Shock.Curr Treat Options Cardiovasc Med. 2016 Jan;18(1):6. doi: 10.1007/s11936-015-0426-6. Curr Treat Options Cardiovasc Med. 2016. PMID: 26758053
-
Interventricular dyssynchrony during continuous-flow left ventricular assist device support: observation using the conductance method.J Artif Organs. 2019 Dec;22(4):348-352. doi: 10.1007/s10047-019-01115-5. Epub 2019 Jul 2. J Artif Organs. 2019. PMID: 31267351
-
Utilization of Percutaneous Mechanical Circulatory Support Devices in Cardiogenic Shock Complicating Acute Myocardial Infarction and High-Risk Percutaneous Coronary Interventions.J Clin Med. 2019 Aug 13;8(8):1209. doi: 10.3390/jcm8081209. J Clin Med. 2019. PMID: 31412669 Free PMC article. Review.
-
Advanced Percutaneous Mechanical Circulatory Support Devices for Cardiogenic Shock.Crit Care Med. 2017 Nov;45(11):1922-1929. doi: 10.1097/CCM.0000000000002676. Crit Care Med. 2017. PMID: 28857849 Free PMC article. Review.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical