Amperometric detection of quantal catecholamine secretion from individual cells on micromachined silicon chips
- PMID: 12585478
- DOI: 10.1021/ac025802m
Amperometric detection of quantal catecholamine secretion from individual cells on micromachined silicon chips
Abstract
We have fabricated electrochemical electrodes in picolitersized wells for measuring catecholamine release from individual cells with millisecond resolution. Each well-electrode roughly conforms to the shape of the cell in order to capture a large fraction of released catecholamine with high time resolution. Using this device, we can resolve spikes in amperometric current corresponding to quantal catecholamine release via exocytosis. In addition, we have combined amperometric recording on the chip with patch-clamp recordings of membrane capacitance as an assay of exocytosis. A quantitative comparison of the two methods suggests that a large fraction of catecholamine release is oxidized on the surface of the well-electrode. This technology has applications in cell-based biosensor development, high-throughput screening of drugs, and basic science investigations.
Similar articles
-
Microwell device for targeting single cells to electrochemical microelectrodes for high-throughput amperometric detection of quantal exocytosis.Anal Chem. 2011 Apr 1;83(7):2445-51. doi: 10.1021/ac1033616. Epub 2011 Feb 28. Anal Chem. 2011. PMID: 21355543 Free PMC article.
-
A microfluidic cell trap device for automated measurement of quantal catecholamine release from cells.Lab Chip. 2009 Dec 7;9(23):3442-6. doi: 10.1039/b913216c. Epub 2009 Sep 30. Lab Chip. 2009. PMID: 19904414
-
Magnetron sputtered diamond-like carbon microelectrodes for on-chip measurement of quantal catecholamine release from cells.Biomed Microdevices. 2008 Oct;10(5):623-9. doi: 10.1007/s10544-008-9173-8. Biomed Microdevices. 2008. PMID: 18493856 Free PMC article.
-
Electrochemical measurement of quantal exocytosis using microchips.Pflugers Arch. 2018 Jan;470(1):97-112. doi: 10.1007/s00424-017-2063-2. Epub 2017 Sep 2. Pflugers Arch. 2018. PMID: 28866728 Free PMC article. Review.
-
Surface-modified CMOS IC electrochemical sensor array targeting single chromaffin cells for highly parallel amperometry measurements.Pflugers Arch. 2018 Jan;470(1):113-123. doi: 10.1007/s00424-017-2067-y. Epub 2017 Sep 9. Pflugers Arch. 2018. PMID: 28889250 Free PMC article. Review.
Cited by
-
A Multilayer MEMS Platform for Single-Cell Electric Impedance Spectroscopy and Electrochemical Analysis.J Microelectromech Syst. 2008 Aug 1;17(4):850-862. doi: 10.1109/JMEMS.2008.921726. J Microelectromech Syst. 2008. PMID: 19756255 Free PMC article.
-
Carbon fiber ultramicrodic electrode electrodeposited with over-oxidized polypyrrole for amperometric detection of vesicular exocytosis from pheochromocytoma cell.Sensors (Basel). 2015 Jan 6;15(1):868-79. doi: 10.3390/s150100868. Sensors (Basel). 2015. PMID: 25569759 Free PMC article.
-
Rapid exocytosis kinetics measured by amperometry within volcano microelectrodes.Analyst. 2023 May 2;148(9):2110-2121. doi: 10.1039/d2an01779b. Analyst. 2023. PMID: 37038889 Free PMC article.
-
Microwell device for targeting single cells to electrochemical microelectrodes for high-throughput amperometric detection of quantal exocytosis.Anal Chem. 2011 Apr 1;83(7):2445-51. doi: 10.1021/ac1033616. Epub 2011 Feb 28. Anal Chem. 2011. PMID: 21355543 Free PMC article.
-
Development and characterization of a diamond-insulated graphitic multi electrode array realized with ion beam lithography.Sensors (Basel). 2014 Dec 30;15(1):515-28. doi: 10.3390/s150100515. Sensors (Basel). 2014. PMID: 25558992 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources