Defects in sphygmomanometers: an important source of error in blood pressure recording
- PMID: 1260385
- PMCID: PMC1639553
- DOI: 10.1136/bmj.1.6014.886
Defects in sphygmomanometers: an important source of error in blood pressure recording
Abstract
Almost half the sphygmomanometers in a teaching hospital group had defects in the control valve which interfered with accurate blood-pressure reading. Ward staff should be taught to check sphygmomanometers regularly and replace control valves; time consumption and cost are low. The cuffs in general use in these hospitals, and of the standard size sold in Britain, had rubber bags which did not encircle the arm of more than half the patients on whom they were used. This deficiency causes over-reading of blood pressure in obese people. The size of error is uncertain but it should be avoided by adopting a cuff with a longer bag.
Similar articles
-
Evaluation of normal and large sphygmomanometer cuffs using the Omron 705CP.J Hum Hypertens. 2001 Feb;15(2):131-4. doi: 10.1038/sj.jhh.1001133. J Hum Hypertens. 2001. PMID: 11317193 Clinical Trial.
-
Rectangular cuffs may overestimate blood pressure in individuals with large conical arms.J Hypertens. 2012 Mar;30(3):530-6. doi: 10.1097/HJH.0b013e32834f98a0. J Hypertens. 2012. PMID: 22278143
-
Influence of cuff size on blood pressure among schoolchildren.Blood Press. 1999;8(3):172-6. doi: 10.1080/080370599439706. Blood Press. 1999. PMID: 10595695
-
The dilemma of choice: which bladder for accurate blood pressure measurement?Z Kardiol. 1996;85 Suppl 3:26-31. Z Kardiol. 1996. PMID: 8896296 Review.
-
Review: a century of confusion; which bladder for accurate blood pressure measurement?J Hum Hypertens. 1996 Sep;10(9):565-72. J Hum Hypertens. 1996. PMID: 8953199 Review.
Cited by
-
Knowledge, attitude, behavior, and practice of interns toward errors in sphygmomanometer and blood pressure measurement.J Clin Hypertens (Greenwich). 2009 Dec;11(12):748-52. doi: 10.1111/j.1751-7176.2009.00204.x. J Clin Hypertens (Greenwich). 2009. PMID: 20021536 Free PMC article. No abstract available.
-
Sphygmomanometers in hospital and family practice: problems and recommendations.Br Med J (Clin Res Ed). 1982 Aug 14;285(6340):469-71. doi: 10.1136/bmj.285.6340.469. Br Med J (Clin Res Ed). 1982. PMID: 6809128 Free PMC article.
-
Sources of inaccuracy in the measurement of adult patients' resting blood pressure in clinical settings: a systematic review.J Hypertens. 2017 Mar;35(3):421-441. doi: 10.1097/HJH.0000000000001197. J Hypertens. 2017. PMID: 27977471 Free PMC article.
-