Normative retardation data corrected for the corneal polarization axis with scanning laser polarimetry
- PMID: 12665235
Normative retardation data corrected for the corneal polarization axis with scanning laser polarimetry
Abstract
Background and objective: To evaluate the distribution of retinal nerve fiber layer thickness measurements in normal eyes corrected for corneal polarization axis (CPA).
Methods: Complete ocular examination, standard automated perimetry, peripapillary and macular scanning laser polarimetry imaging, and CPA measurements were performed in normal eyes. A noninvasive device mounted on a slit lamp that incorporated two crossed linear polarizers and an optical retarder to measure the slow axis of corneal birefringence was constructed. One eye per patient was enrolled. Exclusion criteria consisted of visual acuity of less than 20/40, a refractive error exceeding +/- 5.0 diopter sphere, 2.0 diopter cylinder, or both, previous intraocular surgery, or ocular disease except cataract. According to scanning laser polarimetry, peripapillary retardation parameters within 90% normal limits were recorded as "normal," parameters outside 95% normal limits were recorded as "abnormal," and parameters between 90% and 95% limits were recorded as "borderline."
Results: Fifty-one eyes of 51 subjects (14 men, 37 women) were enrolled (mean age, 51 +/- 17 years). Prediction limits (+/- 2 standard deviation outside of regression line) were calculated for 14 peripapillary retardation parameters. The prediction limits of the average thickness of the retinal nerve fiber layer (microm) were reduced 35% by correction of the CPA; for ellipse average (microm), the limits were reduced 33%. Sixteen of 51 normal eyes (31%) had at least one abnormal retardation parameter, and among these 16 eyes there was a bimodal distribution of CPA values (0 degrees to 20 degrees nasally downward, 60 degrees to 80 degrees nasally downward).
Conclusions: Correction for CPA reduces the variance of normal retardation measurements. Normal eyes with a CPA of 0 degrees to 20 degrees nasally downward and 60 degrees to 80 degrees nasally downward were characterized by "abnormal" summary retardation parameters as interby scanning laser polarimetry.
Similar articles
-
Effect of corneal polarization axis on assessment of retinal nerve fiber layer thickness by scanning laser polarimetry.Am J Ophthalmol. 2000 Jun;129(6):715-22. doi: 10.1016/s0002-9394(00)00353-6. Am J Ophthalmol. 2000. PMID: 10926978
-
Correction for corneal polarization axis improves the discriminating power of scanning laser polarimetry.Am J Ophthalmol. 2002 Jul;134(1):27-33. doi: 10.1016/s0002-9394(02)01512-x. Am J Ophthalmol. 2002. PMID: 12095804
-
Individualized compensation of anterior segment birefringence during scanning laser polarimetry.Invest Ophthalmol Vis Sci. 2002 Jul;43(7):2221-8. Invest Ophthalmol Vis Sci. 2002. PMID: 12091420
-
Stability of corneal polarization axis measurements for scanning laser polarimetry.Ophthalmology. 2001 Jun;108(6):1065-9. doi: 10.1016/s0161-6420(01)00569-3. Ophthalmology. 2001. PMID: 11382630
-
New developments in scanning laser polarimetry for glaucoma.Curr Opin Ophthalmol. 2008 Mar;19(2):136-40. doi: 10.1097/ICU.0b013e3282f4b1c2. Curr Opin Ophthalmol. 2008. PMID: 18301287 Review.
Cited by
-
Detection of progressive retinal nerve fiber layer loss in glaucoma using scanning laser polarimetry with variable corneal compensation.Invest Ophthalmol Vis Sci. 2009 Apr;50(4):1675-81. doi: 10.1167/iovs.08-2712. Epub 2008 Nov 21. Invest Ophthalmol Vis Sci. 2009. PMID: 19029038 Free PMC article.
-
Impact of atypical retardation patterns on detection of glaucoma progression using the GDx with variable corneal compensation.Am J Ophthalmol. 2009 Jul;148(1):155-63.e1. doi: 10.1016/j.ajo.2009.01.021. Epub 2009 Apr 17. Am J Ophthalmol. 2009. PMID: 19375062 Free PMC article.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources