Ventricular oversensing: a study of 101 patients implanted with dual chamber defibrillators and two different lead systems
- PMID: 12685142
- DOI: 10.1046/j.1460-9592.2003.00152.x
Ventricular oversensing: a study of 101 patients implanted with dual chamber defibrillators and two different lead systems
Abstract
Modern dual chamber ICD systems are able to overcome various sensing problems. However, improvement of their performance is still required. The aim of this study was to assess the sensing function in 101 consecutive patients (84 men, 17 women; mean age 63 +/- 12 years; mean follow-up 24 +/- 4 months) implanted with dual chamber defibrillators and integrated (IB) or dedicated bipolar (DB) lead systems. Follow-up data were analyzed for the presence of ventricular oversensing. Oversensing occurred in 25 (25%) patients, significantly more frequent in patients implanted with IB compared to DB lead systems (21/52 vs 4/49, P = 0.0002). Patients with cardiomyopathies (CMs) were more prone to sensing malfunctions than patients with no CM (12/30 vs 13/71, P = 0.04). T wave oversensing (n = 14), respirophasic ventricular oversensing (n = 4), and P wave oversensing (n = 6) were the most common pitfalls of ventricular sensing. P wave oversensing was unique to the IB lead system. CT scans performed in these patients disclosed the position of the RV coil to be proximal to the tricuspid area. Four patients received inappropriate ICD shocks due to oversensing. In all but two patients who received lead revision, oversensing was resolved by noninvasive means.
In conclusion: (1) ventricular oversensing is a common problem occurring in up to 25% of patients with dual chamber ICDs; (2) P wave oversensing is a ventricular sensing problem affecting function of 11% of dual chamber devices with IB lead systems; (3) IB leads are significantly more susceptible to T wave and P wave oversensing than DB leads; and (4) patients with cardiomyopathies are more prone to oversensing than patients with other heart diseases.
Similar articles
-
T-wave oversensing in patients with Brugada syndrome: true bipolar versus integrated bipolar implantable cardioverter defibrillator leads: multicenter retrospective study.Circ Arrhythm Electrophysiol. 2015 Aug;8(4):792-8. doi: 10.1161/CIRCEP.115.002871. Epub 2015 Jun 3. Circ Arrhythm Electrophysiol. 2015. PMID: 26041387
-
Prevalence of sensing abnormalities in dual chamber implantable cardioverter defibrillators.Ann Noninvasive Electrocardiol. 2003 Jul;8(3):219-26. doi: 10.1046/j.1542-474x.2003.08309.x. Ann Noninvasive Electrocardiol. 2003. PMID: 14510657 Free PMC article.
-
Potential role of home monitoring to reduce inappropriate shocks in implantable cardioverter-defibrillator patients due to lead failure.Europace. 2009 Apr;11(4):483-8. doi: 10.1093/europace/eun350. Epub 2008 Dec 22. Europace. 2009. PMID: 19103654
-
Sensing and detection in Medtronic implantable cardioverter defibrillators.Herzschrittmacherther Elektrophysiol. 2016 Sep;27(3):193-212. doi: 10.1007/s00399-016-0450-6. Epub 2016 Sep 8. Herzschrittmacherther Elektrophysiol. 2016. PMID: 27624809 Review. English.
-
Inappropriate pacing due to autoperpetuation of the ventricular rate stabilization algorithm: a manifestation of T wave oversensing by ICDs.Pacing Clin Electrophysiol. 2000 Sep;23(9):1446-7. doi: 10.1111/j.1540-8159.2000.tb00980.x. Pacing Clin Electrophysiol. 2000. PMID: 11025907 Review.
Cited by
-
R-wave double counting secondary to antiarrhythmics and ablation leads to unnecessary shock.HeartRhythm Case Rep. 2020 Jul 31;6(10):782-785. doi: 10.1016/j.hrcr.2020.07.018. eCollection 2020 Oct. HeartRhythm Case Rep. 2020. PMID: 33101955 Free PMC article. No abstract available.
-
An inappropriately-appropriate ICD therapy: What is the mechanism and what interventions can be done?Clin Case Rep. 2023 Oct 23;11(10):e8082. doi: 10.1002/ccr3.8082. eCollection 2023 Oct. Clin Case Rep. 2023. PMID: 37881196 Free PMC article.
-
Unmasking of myopotential oversensing by an integrated bipolar defibrillator lead following AV node ablation.Herzschrittmacherther Elektrophysiol. 2012 Jun;23(2):131-4. doi: 10.1007/s00399-012-0172-3. Herzschrittmacherther Elektrophysiol. 2012. PMID: 22457167
-
Transient hyperkalemia-induced T-wave oversensing as a cause of inappropriate ICD discharges.J Cardiol Cases. 2013 Oct 2;8(5):164-167. doi: 10.1016/j.jccase.2013.07.008. eCollection 2013 Nov. J Cardiol Cases. 2013. PMID: 30534283 Free PMC article.
-
Treatment of symptomatic bradycardia due to T-wave oversensing with implantation of a new generator incorporating delayed decay and threshold start sensitization algorithms.HeartRhythm Case Rep. 2020 Jul 1;6(12):891-895. doi: 10.1016/j.hrcr.2020.06.024. eCollection 2020 Dec. HeartRhythm Case Rep. 2020. PMID: 33365232 Free PMC article. No abstract available.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous