Alveolar pressure monitoring: an evaluation in a lung model and in patients with acute lung injury

Affiliations
Comparative Study

Alveolar pressure monitoring: an evaluation in a lung model and in patients with acute lung injury

S Sondergaard et al. Intensive Care Med. 2003 Jun.

Abstract

Objectives: We evaluated an algorithm for continuous on-line monitoring of alveolar pressure over time in a lung model with lower and upper inflection points and variable resistance ratios and in patients with acute lung injury. The algorithm is based on "static" pressure/volume curves obtained from tracheal pressure measurements under dynamic conditions.

Design and setting: Experimental and clinical evaluation of algorithm in a university hospital laboratory and intensive care unit.

Patients: Ten patients undergoing postoperative respiratory therapy (feasibility of tracheal measurement) and ten patients with acute lung injury undergoing ventilator treatment (evaluation of algorithm).

Measurements and results: Direct tracheal pressure measurements with a catheter inserted through the endotracheal tube. Comparison of measured alveolar and the dynostatic alveolar pressure vs. time in a lung model with changes in five ventilatory parameters. Examples of clinical monitoring are reported. In the model there was excellent agreement between alveolar pressures obtained by the algorithm, the dynostatic alveolar pressure, and measured alveolar pressure at all ventilator settings. For inspiratory/expiratory resistance ratios between 1:2.1-2.1:1, the dynostatic alveolar pressure was within +/-1.5 cm H(2)O of measured alveolar pressure. In patients the technique for direct tracheal pressure measurement using a catheter inserted through the endotracheal tube functioned satisfactorily with intermittent air flushes for cleansing.

Conclusions: Using a thin tracheal pressure catheter inserted through the endotracheal tube alveolar pressure allows continuous bedside monitoring with ease and precision using the dynostatic algorithm. The method is unaffected by tube and connector geometry or by secretions.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Acta Anaesthesiol Scand. 2001 Sep;45(8):1046-50 - PubMed
    1. Intensive Care Med. 1999 Oct;25(10):1040-3 - PubMed
    1. Am J Respir Crit Care Med. 1998 Dec;158(6):1831-8 - PubMed
    1. Eur Respir J. 1996 Feb;9(2):262-73 - PubMed
    1. Acta Anaesthesiol Scand. 2001 Feb;45(2):173-9 - PubMed

Publication types

MeSH terms

LinkOut - more resources