Validation of a tonometric noninvasive arterial blood pressure monitor in the intensive care setting
- PMID: 12694001
- DOI: 10.1046/j.1365-2044.2003.03122.x
Validation of a tonometric noninvasive arterial blood pressure monitor in the intensive care setting
Abstract
Intra-arterial measurement is considered the gold standard for continuous, beat-to-beat arterial blood pressure monitoring. However, arterial cannulation can be difficult and may cause complications such as thrombosis and ischaemia. Recently, a tonometric system, the Colin CBM-7000 has been developed for noninvasive beat-to-beat measurement of arterial blood pressure from the radial artery. We assessed the level of agreement between the CBM-7000 and invasive radial artery measurements in 15 patients on a neuro-intensive care unit. Agreement of systolic, diastolic and mean arterial pressure values was limited, with approximately 34% of mean arterial pressures differing by over 10 mmHg. In many cases, this was due to a downward drift of the noninvasive measurements over time. Furthermore, there was a tendency to underestimate low pressures and overestimate high pressures. In our opinion, the Colin CBM-7000 cannot be recommended for continuous blood pressure monitoring in the intensive care setting.
Similar articles
-
Beat-to-beat agreement of noninvasive tonometric and intra-radial arterial blood pressure during microgravity and hypergravity generated by parabolic flights.Blood Press Monit. 2007 Dec;12(6):357-62. doi: 10.1097/01.mbp.0000209088.78235.59. Blood Press Monit. 2007. PMID: 18004103
-
Non-invasive continuous arterial pressure measurement based on radial artery tonometry in the intensive care unit: a method comparison study using the T-Line TL-200pro device.Br J Anaesth. 2013 Aug;111(2):185-90. doi: 10.1093/bja/aet025. Epub 2013 Mar 13. Br J Anaesth. 2013. PMID: 23491946
-
Comparison of non-invasive blood pressure monitoring using modified arterial applanation tonometry with intra-arterial measurement.J Clin Monit Comput. 2018 Feb;32(1):13-22. doi: 10.1007/s10877-017-9984-3. Epub 2017 Jan 19. J Clin Monit Comput. 2018. PMID: 28105538 Clinical Trial.
-
Blood pressure measurement during exercise: a review.Med Sci Sports Exerc. 1997 Jan;29(1):149-59. doi: 10.1097/00005768-199701000-00022. Med Sci Sports Exerc. 1997. PMID: 9000169 Review.
-
Continuous intra-arterial blood gas monitoring.Intensive Care Med. 1996 Aug;22(8):818-28. doi: 10.1007/BF01709527. Intensive Care Med. 1996. PMID: 8880253 Review.
Cited by
-
The relationship between orthostatic hypotension and falling in older adults.Clin Auton Res. 2014 Feb;24(1):3-13. doi: 10.1007/s10286-013-0219-5. Epub 2013 Nov 20. Clin Auton Res. 2014. PMID: 24253897 Review.
-
Noninvasive continuous hemodynamic monitoring.J Clin Monit Comput. 2012 Aug;26(4):267-78. doi: 10.1007/s10877-012-9375-8. Epub 2012 Jun 14. J Clin Monit Comput. 2012. PMID: 22695821 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical