Relationship between magnetic field strength and magnetic-resonance-related acoustic noise levels
- PMID: 12695886
- DOI: 10.1007/s10334-003-0005-9
Relationship between magnetic field strength and magnetic-resonance-related acoustic noise levels
Abstract
The need for better signal-to-noise ratios and resolution has pushed magnetic resonance imaging (MRI) towards high-field MR-scanners for which only little data on MR-related acoustic noise production have been published. The purpose of this study was to validate the theoretical relationship of sound pressure level (SPL) and static magnetic field strength. This is relevant for allowing adequate comparisons of acoustic data of MR systems at various magnetic field strengths. Acoustic data were acquired during various pulse sequences at field strengths of 0.5, 1.0, 1.5 and 2.0 Tesla using the same MRI unit by means of a Helicon rampable magnet. Continuous-equivalent, i.e. time-averaged, linear SPLs and 1/3-octave band frequencies were recorded. Ramping from 0.5 to 1.0 Tesla and from 1.0 to 2.0 Tesla resulted in an SPL increase of 5.7 and 5.2 dB(L), respectively, when averaged over the various pulse sequences. Most of the acoustic energy was in the 1-kHz frequency band, irrespective of magnetic field strength. The relation between field strength and SPL was slightly non-linear, i.e. a slightly less increase at higher field strengths, presumably caused by the elastic properties of the gradient coil encasings.
Similar articles
-
Efficacy of passive acoustic screening: implications for the design of imager and MR-suite.J Magn Reson Imaging. 2003 Feb;17(2):270-5. doi: 10.1002/jmri.10251. J Magn Reson Imaging. 2003. PMID: 12541235
-
Evaluation of magnetic resonance imaging acoustic noise reduction technology by magnetic gradient waveform control.Magn Reson Imaging. 2019 Nov;63:170-177. doi: 10.1016/j.mri.2019.08.015. Epub 2019 Aug 16. Magn Reson Imaging. 2019. PMID: 31425801
-
Simulation study of noise reduction methods for a split MRI system using a finite element method.Med Phys. 2015 Dec;42(12):7122-31. doi: 10.1118/1.4935864. Med Phys. 2015. PMID: 26632066
-
Body MR Imaging: Artifacts, k-Space, and Solutions.Radiographics. 2015 Sep-Oct;35(5):1439-60. doi: 10.1148/rg.2015140289. Epub 2015 Jul 24. Radiographics. 2015. PMID: 26207581 Free PMC article. Review.
-
[Methodological aspects of functional neuroimaging at high field strength: a critical review].Rofo. 2007 Sep;179(9):925-31. doi: 10.1055/s-2007-963195. Epub 2007 Jun 26. Rofo. 2007. PMID: 17594628 Review. German.
Cited by
-
Gradient and shim technologies for ultra high field MRI.Neuroimage. 2018 Mar;168:59-70. doi: 10.1016/j.neuroimage.2016.11.033. Epub 2016 Nov 30. Neuroimage. 2018. PMID: 27915120 Free PMC article. Review.
-
Sustainable low-field cardiovascular magnetic resonance in changing healthcare systems.Eur Heart J Cardiovasc Imaging. 2022 Jun 1;23(6):e246-e260. doi: 10.1093/ehjci/jeab286. Eur Heart J Cardiovasc Imaging. 2022. PMID: 35157038 Free PMC article. Review.
-
Acoustic noise concerns in functional magnetic resonance imaging.Hum Brain Mapp. 2003 Nov;20(3):123-41. doi: 10.1002/hbm.10134. Hum Brain Mapp. 2003. PMID: 14601139 Free PMC article. Review.
-
Silent zero TE MR neuroimaging: Current state-of-the-art and future directions.Prog Nucl Magn Reson Spectrosc. 2021 Apr;123:73-93. doi: 10.1016/j.pnmrs.2021.03.002. Epub 2021 Mar 26. Prog Nucl Magn Reson Spectrosc. 2021. PMID: 34078538 Free PMC article. Review.
-
Methodological challenges and solutions in auditory functional magnetic resonance imaging.Front Neurosci. 2014 Aug 21;8:253. doi: 10.3389/fnins.2014.00253. eCollection 2014. Front Neurosci. 2014. PMID: 25191218 Free PMC article. Review.