Evaluation of the clinical applicability of bromcresol purple method for measuring serum albumin
- PMID: 12697479
Evaluation of the clinical applicability of bromcresol purple method for measuring serum albumin
Abstract
Objective: To evaluate the clinical applicability of bromcresol purple (BCP) method in serum albumin measurement, and to investigate the possibility to render the results thus obtained in agreement with those by traditional bromcresol green (BCG) method.
Methods: The assessment of BCP method was conducted in comparison with BCG method by measuring their linearity ranges, coefficients of variation (CV) in batch and between batches, recovery rates, and resistance to interference by certain normal clinical factors.
Results: The linearity of BCP method was 10 to 80 g/L, CV was 0.95% in batch and 3.03% between batches, with an average recovery rate of 99.4%, showing no obvious interference by 252 micromol/L bilirubin, 4% Trilitid, or by 4 g/L hemoglobin separately. Close correlation was observed between the measurement results by the 2 methods (y=1.08x+5.98). The adult reference range was 34-49 g/L by BCP method, which was 6 g/L lower, on average, than that by traditional BCG method.
Conclusion: BCP method is more specific than BCG method in serum albumin measurement, and the results of the former can be easily and accurately converted so as to be comparable with the results by the latter method.
Similar articles
-
[Controversies in the determination of serum albumin concentration in chronic liver diseases].Rinsho Byori. 2006 Oct;54(10):1008-12. Rinsho Byori. 2006. PMID: 17133989 Japanese.
-
The impact of the assay for measuring albumin on corrected ('adjusted') calcium concentrations.Nephrol Dial Transplant. 2009 Jun;24(6):1834-8. doi: 10.1093/ndt/gfn747. Epub 2009 Jan 31. Nephrol Dial Transplant. 2009. PMID: 19182240
-
The impact of change in albumin assay on reference intervals, prevalence of 'hypoalbuminaemia' and albumin prescriptions.Ann Clin Biochem. 2016 Jan;53(Pt 1):112-6. doi: 10.1177/0004563215599560. Epub 2015 Jul 20. Ann Clin Biochem. 2016. PMID: 26195486
-
Serum and urine albumin: a progress report on their measurement and clinical significance.Clin Chim Acta. 1997 Feb 3;258(1):3-20. doi: 10.1016/s0009-8981(96)06446-7. Clin Chim Acta. 1997. PMID: 9049439 Review.
-
The measurement of albumin in serum and plasma.Ann Clin Biochem. 1985 Nov;22 ( Pt 6):565-78. doi: 10.1177/000456328502200604. Ann Clin Biochem. 1985. PMID: 3935035 Review. No abstract available.