Measurement of cardiac output after cardiac surgery by a new transesophageal Doppler device
- PMID: 12698405
- DOI: 10.1053/jcan.2003.50
Measurement of cardiac output after cardiac surgery by a new transesophageal Doppler device
Abstract
Objectives: Assessment of hemodynamics by transesophageal Doppler devices (TDD) may be a less invasive alternative to the pulmonary artery catheter. In contrast to the TDD evaluated so far, a new monitor (HemoSonic100) measures both blood flow velocity and the diameter of the descending aorta. The aim of this study was to assess the accuracy of the cardiac output/index (CO/CI) measured by this device compared with the CO/CI measured by thermodilution.
Design: Prospective nonrandomized study.
Setting: Community hospital; university-based statistician.
Participants: Twenty-two patients.
Interventions: Elective coronary artery bypass grafting and/or valve replacement/repair.
Measurements and main results: After routine cardiac surgery, CO/CI was determined in the intensive care unit by iced-water bolus (IWB), continuous cardiac index (CCI) assessment, and the TDD. Matched measurements were made with each patient at intervals of 30 minutes. Six percent of sets were incomplete because of failed signal detection by the TDD. Bland-Altman analysis revealed a mean bias of 0.23 L/min/m(2) for TDD and IWB. Mean bias for CCI and IWB was 0.11 L/min/m(2). The correlation between TDD and IWB (r(2) = 0.09) for cardiac index was found to be inferior to the correlation between CCI and IWB (r(2) = 0.65). Trend analysis between sequential measurements (T1-4: dTDD, dCCI, dIWB) showed a lower correlation between dTDD and dIWB (r(2) = 0.1) compared with the correlation between dCCI and dIWB (r(2) = 0.44).
Conclusion: The transesophageal Doppler device (HemoSonic100) cannot be recommended as a sole method for monitoring cardiac output in patients after cardiac surgery.
Copyright 2003 Elsevier Inc. All rights reserved.
Similar articles
-
Pulsed dye densitometry with two different sensor types for cardiac output measurement after cardiac surgery: a comparison with the thermodilution technique.Acta Anaesthesiol Scand. 2004 May;48(5):653-7. doi: 10.1111/j.1399-6576.2004.00371.x. Acta Anaesthesiol Scand. 2004. PMID: 15101865 Clinical Trial.
-
Accuracy of cardiac output measurements with pulse contour analysis (PulseCO) and Doppler echocardiography during off-pump coronary artery bypass grafting.Eur J Anaesthesiol. 2008 Mar;25(3):243-8. doi: 10.1017/S0265021507002979. Epub 2007 Nov 9. Eur J Anaesthesiol. 2008. PMID: 17996125
-
Cardiac output measurement during infrarenal aortic surgery: echo-esophageal Doppler versus thermodilution catheter.J Cardiothorac Vasc Anesth. 2006 Feb;20(1):26-30. doi: 10.1053/j.jvca.2005.07.029. J Cardiothorac Vasc Anesth. 2006. PMID: 16458209
-
Utility of esophageal Doppler as a minimally invasive hemodynamic monitor: a review.Can J Anaesth. 2002 Apr;49(4):393-401. doi: 10.1007/BF03017329. Can J Anaesth. 2002. PMID: 11927480 Review.
-
A meta-analysis of the accuracy and precision of the ultrasonic cardiac output monitor (USCOM).Anaesthesia. 2012 Nov;67(11):1266-71. doi: 10.1111/j.1365-2044.2012.07311.x. Epub 2012 Aug 29. Anaesthesia. 2012. PMID: 22928650 Review.
Cited by
-
Methods in pharmacology: measurement of cardiac output.Br J Clin Pharmacol. 2011 Mar;71(3):316-30. doi: 10.1111/j.1365-2125.2010.03798.x. Br J Clin Pharmacol. 2011. PMID: 21284692 Free PMC article. Review.
-
Transesophageal Doppler devices: A technical review.J Clin Monit Comput. 2009 Dec;23(6):391-401. doi: 10.1007/s10877-009-9204-x. Epub 2009 Oct 20. J Clin Monit Comput. 2009. PMID: 19842049 Review.
-
The validity of trans-esophageal Doppler ultrasonography as a measure of cardiac output in critically ill adults.Intensive Care Med. 2004 Nov;30(11):2060-6. doi: 10.1007/s00134-004-2430-2. Epub 2004 Sep 11. Intensive Care Med. 2004. PMID: 15368037 Review.
-
[Measurement of cardiac output].Anaesthesist. 2005 Nov;54(11):1135-51; quiz 1152-3. doi: 10.1007/s00101-005-0924-0. Anaesthesist. 2005. PMID: 16228152 Review. German.
-
Goal-directed intraoperative therapy based on autocalibrated arterial pressure waveform analysis reduces hospital stay in high-risk surgical patients: a randomized, controlled trial.Crit Care. 2010;14(1):R18. doi: 10.1186/cc8875. Epub 2010 Feb 15. Crit Care. 2010. PMID: 20156348 Free PMC article. Clinical Trial.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials