High-throughput flow cytometry: validation in microvolume bioassays
- PMID: 12701132
- DOI: 10.1002/cyto.a.10035
High-throughput flow cytometry: validation in microvolume bioassays
Abstract
Background: We recently reported an automated sample handling system, designated HyperCyt, by which samples are aspirated from microplate wells and delivered to the flow cytometer for analysis at rates approaching 100 samples per minute. In this approach, an autosampler and peristaltic pump introduce samples into a tubing line that directly connects to the flow cytometer. Air bubbles are inserted between samples to prevent sample dispersion. In the present work, we compare results of HyperCyt with those of conventional manual flow cytometric analysis in representative flow cytometric bioassays and describe a cell suspension method in which HyperCyt exploits the use of microvolume wells.
Methods: Human eosinophils and neutrophils were treated with trypsin to generate a wide (>25-fold) range of membrane P-selectin glycoprotein ligand-1 (PSGL-1) expression and then stained with fluorescent anti-PSGL-1 antibodies. Human peripheral blood mononuclear cells were stained with fluorescein isothiocyanate- and phycoerythrin-conjugated monoclonal antibodies for multiparameter immunophenotype analysis. U937 cells labeled with PKH62GL were used to assess cell settling in microplate wells.
Results: Differences in PSGL-1 expression levels were detected by HyperCyt autosampling of leukocytes from 96-well plates at an analysis rate of approximately 1.5 s/well. HyperCyt measurements linearly correlated with parallel manual measurements (r(2) = 0.98). Lymphocyte subpopulations were accurately distinguished and reproducibly quantified in multiparameter immunophenotyping assays performed over a range of HyperCyt analysis rates (1.4-5.5 s/sample). When assay volumes were reduced to 10 microl/well in 60-well Terasaki plates, cells could be maintained in uniform suspension for up to 30 min by periodically inverting plates on a rotating carousel before HyperCyt analysis. HyperCyt analysis of five fluorescence-level Cyto-Plex beads sampled from Terasaki plate microwells at 2.5 s/well produced highly reproducible results over a wide range of input bead concentrations (from 7 x 10(5) to 20 x 10(6) beads/ml) that linearly correlated with manual analysis results.
Conclusions: The HyperCyt autosampling system enabled a 10-fold or greater increase in sample throughput compared with conventional manual flow cytometric sample analysis, with comparable analysis results. Assays were performed efficiently in 10-microl volumes to enable significant reagent cost savings, use of quantity-limited reagents at otherwise prohibitive concentrations, and maintenance of uniform suspensions of cells for prolonged periods.
Copyright 2003 Wiley-Liss, Inc.
Similar articles
-
High throughput flow cytometry.Cytometry. 2001 May 1;44(1):83-90. Cytometry. 2001. PMID: 11309812
-
High-throughput screening with HyperCyt flow cytometry to detect small molecule formylpeptide receptor ligands.J Biomol Screen. 2005 Jun;10(4):374-82. doi: 10.1177/1087057105274532. J Biomol Screen. 2005. PMID: 15964939
-
Fluorescence-intensity multiplexing: simultaneous seven-marker, two-color immunophenotyping using flow cytometry.Cytometry A. 2004 Oct;61(2):142-52. doi: 10.1002/cyto.a.20037. Cytometry A. 2004. PMID: 15382027
-
Clinical applications of flow cytometry.Clin Lab Sci. 1993 May-Jun;6(3):174-6. Clin Lab Sci. 1993. PMID: 10146218 Review.
-
Flow cytometry for high-throughput, high-content screening.Curr Opin Chem Biol. 2004 Aug;8(4):392-8. doi: 10.1016/j.cbpa.2004.06.007. Curr Opin Chem Biol. 2004. PMID: 15288249 Review.
Cited by
-
High-content screening: flow cytometry analysis.Methods Mol Biol. 2009;486:151-65. doi: 10.1007/978-1-60327-545-3_11. Methods Mol Biol. 2009. PMID: 19347622 Free PMC article.
-
Fate Decision Between Group 3 Innate Lymphoid and Conventional NK Cell Lineages by Notch Signaling in Human Circulating Hematopoietic Progenitors.J Immunol. 2017 Oct 15;199(8):2777-2793. doi: 10.4049/jimmunol.1601711. Epub 2017 Sep 11. J Immunol. 2017. PMID: 28893953 Free PMC article.
-
The monoamine oxidase A inhibitor clorgyline is a broad-spectrum inhibitor of fungal ABC and MFS transporter efflux pump activities which reverses the azole resistance of Candida albicans and Candida glabrata clinical isolates.Antimicrob Agents Chemother. 2012 Mar;56(3):1508-15. doi: 10.1128/AAC.05706-11. Epub 2011 Dec 27. Antimicrob Agents Chemother. 2012. PMID: 22203607 Free PMC article.
-
Global phenotypic characterisation of human platelet lysate expanded MSCs by high-throughput flow cytometry.Sci Rep. 2018 Mar 2;8(1):3907. doi: 10.1038/s41598-018-22326-5. Sci Rep. 2018. PMID: 29500387 Free PMC article.
-
A multifunctional androgen receptor screening assay using the high-throughput Hypercyt flow cytometry system.Cytometry A. 2008 May;73(5):390-9. doi: 10.1002/cyto.a.20552. Cytometry A. 2008. PMID: 18340645 Free PMC article.
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources