Revision total hip arthroplasty performed after fracture of a ceramic femoral head. A multicenter survivorship study
- PMID: 12728032
- DOI: 10.2106/00004623-200305000-00009
Revision total hip arthroplasty performed after fracture of a ceramic femoral head. A multicenter survivorship study
Abstract
Background: The alumina ceramic femoral head was introduced for total hip arthroplasty approximately thirty years ago. One of its main drawbacks was the risk of implant fracture. The aim of this study was to examine the results of revision total hip replacement performed specifically to treat a fracture of a ceramic femoral head and to identify technical factors that affected the outcomes.
Methods: One hundred and five surgical revisions to treat a fracture of a ceramic femoral head, performed at thirty-five institutions, were studied. The patients were examined clinically by the operating surgeon at the time of the last follow-up. The surgeon provided the latest follow-up radiographs, which were compared with the immediate postoperative radiographs. The success of the revisions was assessed with Kaplan-Meier survivorship analysis, with the need for repeat revision as the end point. We evaluated the complication rate and radiographic changes indicative of implant loosening. The average duration of follow-up between the index revision and the last clinical and radiographic review was 3.5 years (range, six months to twenty years).
Results: Following the revisions, radiographic evidence of cup loosening was seen in twenty-two hips (21%) and radiographic evidence of femoral loosening was seen in twenty-two (21%). One or several repeat revisions were necessary in thirty-three patients (31%) because of infection (four patients), implant loosening (twenty), osteolysis (eight), or fracture of the revision femoral head component (one). The survival rate at five years was 63% (95% confidence interval, 51% to 75%). The survival rate was significantly worse when the cup had not been changed, when the new femoral head was made of stainless steel, when a total synovectomy had not been done, and when the patient was less than fifty years old.
Conclusions: Fracture of a ceramic femoral head component is a rare but potentially serious event. A suitable surgical approach, including total synovectomy, cup exchange, and insertion of a cobalt-chromium or new ceramic femoral ball minimizes the chance of early loosening of the implants and the need for one or more repeat revisions.
Level of evidence: Therapeutic study, Level IV (case series [no, or historical, control group]).
Similar articles
-
Mid-term results of third-generation alumina-on-alumina ceramic bearings in cementless total hip arthroplasty: a ten-year minimum follow-up.J Bone Joint Surg Am. 2012 Jan 18;94(2):138-44. doi: 10.2106/JBJS.J.00331. J Bone Joint Surg Am. 2012. PMID: 22258000
-
Ceramic-on-ceramic THA Implants in Patients Younger Than 20 Years.Clin Orthop Relat Res. 2016 Feb;474(2):520-7. doi: 10.1007/s11999-015-4546-9. Epub 2015 Sep 4. Clin Orthop Relat Res. 2016. PMID: 26341897 Free PMC article.
-
Impaction allografting with cement for revision of the femoral component. A minimum four-year follow-up study with use of a precoated femoral stem.J Bone Joint Surg Am. 1999 Aug;81(8):1080-92. doi: 10.2106/00004623-199908000-00004. J Bone Joint Surg Am. 1999. PMID: 10466641
-
The Merete BioBall system in hip revision surgery: A systematic review.Orthop Traumatol Surg Res. 2018 Dec;104(8):1171-1178. doi: 10.1016/j.otsr.2018.06.016. Epub 2018 Oct 31. Orthop Traumatol Surg Res. 2018. PMID: 30391216
-
Outcome of total hip arthroplasties with cement. Results after implantation of the Harris Design 2 prosthesis, 1983-1985.J Arthroplasty. 1996 Oct;11(7):789-96. doi: 10.1016/s0883-5403(96)80178-1. J Arthroplasty. 1996. PMID: 8934318 Review.
Cited by
-
Factors affecting modular acetabular ceramic liner insertion: a biomechanical analysis.Clin Orthop Relat Res. 2012 Feb;470(2):402-9. doi: 10.1007/s11999-011-2193-3. Clin Orthop Relat Res. 2012. PMID: 22125246 Free PMC article.
-
Ceramic on crosslinked polyethylene in total hip replacement: any better than metal on crosslinked polyethylene?Iowa Orthop J. 2009;29:1-4. Iowa Orthop J. 2009. PMID: 19742076 Free PMC article.
-
Liner dissociation leading to catastrophic failure of an Oxinium femoral head.Arthroplast Today. 2018 Nov 28;5(1):21-25. doi: 10.1016/j.artd.2018.09.009. eCollection 2019 Mar. Arthroplast Today. 2018. PMID: 31020016 Free PMC article.
-
Ceramic liner fracture and impingement in total hip arthroplasty.HSS J. 2015 Feb;11(1):50-5. doi: 10.1007/s11420-014-9415-7. Epub 2014 Oct 21. HSS J. 2015. PMID: 25737669 Free PMC article.
-
Head material influences survival of a cemented total hip prosthesis in the Norwegian Arthroplasty Register.Clin Orthop Relat Res. 2012 Nov;470(11):3007-13. doi: 10.1007/s11999-012-2396-2. Clin Orthop Relat Res. 2012. PMID: 22644422 Free PMC article.